Quantification of breast density with dual energy mammography: An experimental feasibility study
Authors:
- 26 January 2010
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 37 (2) , 793-801
Abstract
Methods: The dual energy mammography system used a tungsten anode x-ray tube with a 50 mu m rhodium beam filter for low energy images and a 300 mu m copper beam filter for high energy images. Glandular and adipose equivalent phantoms of uniform thickness were used to calibrate a dual energy basis decomposition algorithm. Four different phantom studies were used to evaluate the technique. The first study consisted of phantoms with thicknesses of 2.5-8.5 cm in 0.5 cm steps with variable densities centered at a mean of 28%. The second study consisted of phantoms at a fixed thickness of 4.0 cm, which ranged in densities from 0% to 100% in increments of 12.5%. The third study consisted of 4.0 cm thick phantoms at densities of 25%, 50% and 75% each imaged at three areal sizes, approximately 62.5, 125, and 250 cm(2), in order to assess the effect of breast size on density measurement. The fourth study consisted of step phantoms designed to more closely mimic the shape of a female breast with maximal thicknesses from 3.0 to 7.0 cm at a fixed density of 50%. All images were corrected for x-ray scatter. Results: The RMS errors in breast density measurements were 0.44% for the variable thickness phantoms, 0.64% for the variable density phantoms, 2.87% for the phantoms of different areal sizes, and 4.63% for step phantoms designed to closely resemble the shape of a breast. Conclusions: The results of the phantom studies indicate that dual energy mammography can be used to measure breast density with an RMS error of approximately 5%.Keywords
Funding Information
- National Cancer Institute (R01 CA136871)
This publication has 41 references indexed in Scilit:
- Scatter correction in digital mammography based on image deconvolutionPhysics in Medicine & Biology, 2010
- Mammographic Density and the Risk and Detection of Breast CancerNew England Journal of Medicine, 2007
- Dual-energy mammography: simulation studiesPhysics in Medicine & Biology, 2006
- Evaluation of the minimum iodine concentration for contrast-enhanced subtraction mammographyPhysics in Medicine & Biology, 2006
- Breast composition measurements using retrospective standard mammogram form (SMF)Physics in Medicine & Biology, 2006
- Evaluation of dual-energy subtraction of digital mammography images under conditions found in a commercial unitPhysics in Medicine & Biology, 2006
- A representation for mammographic image processingMedical Image Analysis, 1996
- Dual-energy subtraction mammographyJournal of Digital Imaging, 1995
- The quantitative analysis of mammographic densitiesPhysics in Medicine & Biology, 1994
- Ein Programm für die Berechnung von diagnostischen RöntgenspektrenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1985