Simple flow system for the rapid pre-concentration and potentiometric determination of fluoride using a micro-column and a wall-jet electrode

Authors:
  • YOYoshito Okabayashi
    Yoshito Okabayashi
  • MHMichiyo Hikawa
    Michiyo Hikawa
  • TNTerumichi Nakagawa
    Terumichi Nakagawa
  • HTHisashi Tanaka
    Hisashi Tanaka
  • MCMasahiko Chikuma
    Masahiko Chikuma

Abstract
A flow system for the selective pre-concentration and determination of p.p.b. levels of fluoride has been developed. The system consisted of two micro-columns (150 × 0.5 mm i.d.) packed with an anion-exchange resin loaded with alizarin fluorine blue sulphonate-lanthanum complex and a fluoride ion-selective electrode in a wall-jet type cell. The fluoride present in 1 ml of sample was adsorbed on the resin and eluted with 1 M sodium hydroxide containing 0.5 M sodium chloride. This system allowed 24 samples to be treated in 1 h. Interference by the ubiquitous foreign ions Ca2+, Mg2+ and H2PO4 – at levels 300 times higher than that of fluoride was not observed but Al3+ and Fe3+ interfered seriously at much lower levels; this interference could be eliminated by application of a previously reported method.