A randomised open-label cross-over study of inhaler errors, preference and time to achieve correct inhaler use in patients with COPD or asthma: comparison of ELLIPTA with other inhaler devices

Authors:

Abstract
Errors in the use of different inhalers were investigated in patients naive to the devices under investigation in a multicentre, single-visit, randomised, open-label, cross-over study. Patients with chronic obstructive pulmonary disease (COPD) or asthma were assigned to ELLIPTA vs DISKUS (Accuhaler), metered-dose inhaler (MDI) or Turbuhaler. Patients with COPD were also assigned to ELLIPTA vs Handihaler or Breezhaler. Patients demonstrated inhaler use after reading the patient information leaflet (PIL). A trained investigator assessed critical errors (i.e., those likely to result in the inhalation of significantly reduced, minimal or no medication). If the patient made errors, the investigator demonstrated the correct use of the inhaler, and the patient demonstrated inhaler use again. Fewer COPD patients made critical errors with ELLIPTA after reading the PIL vs: DISKUS, 9/171 (5%) vs 75/171 (44%); MDI, 10/80 (13%) vs 48/80 (60%); Turbuhaler, 8/100 (8%) vs 44/100 (44%); Handihaler, 17/118 (14%) vs 57/118 (48%); Breezhaler, 13/98 (13%) vs 45/98 (46%; all PP=0.221); MDI (2/32 (6%) vs 8/32 (25%), P=0.074) and significantly fewer vs Turbuhaler (3/60 (5%) vs 20/60 (33%), PP⩽0.002). Significantly, fewer COPD patients using ELLIPTA made critical errors after reading the PIL vs other inhalers. More asthma and COPD patients preferred ELLIPTA over comparator inhalers.

This publication has 42 references indexed in Scilit: