Microwave digestion preparation and ICP determination of boron in human plasma
Authors:
- 1 April 1993
- journal article
- Published by Springer Nature in Biological Trace Element Research
- Vol. 37 (1) , 17-25
Abstract
A microwave digestion procedure, followed by Inductively Coupled Argon Plasma Spectroscopy, is described for the determination of boron (B) in human plasma. The National Institute of Standards and Technology (NIST) currently does not certify the concentration of B in any substance. The NIST citrus leaves 1572 (CL) Standard Reference Material (SRM) and wheat flour 1567a (WF) were chosen to determine the efficacy of digestion. CL and WF values compare favorably to those obtained from an open-vessel, wet digestion followed by ICP, and by neutron activation and mass spectrometric measurements. Plasma samples were oxidized by doubled-distilled ultrapure HNO3 in 120 mL PFA Teflon™ vessels. An MDS-81D microwave digestion procedure allows for rapid and relatively precise determination of B in human plasma, while limiting handling hazards and sources of contamination.Keywords
This publication has 3 references indexed in Scilit:
- Dietary boron modified the effects of magnesium and molybdenum on mineral metabolism in the cholecalciferol-deficient chickBiological Trace Element Research, 1989
- Analysis of ultratrace lithium and boron by neutron activation and mass-spectrometric measurement of 3He and 4HeApplied Radiation and Isotopes, 1987
- Boron, Cadmium, Chromium, and Nickel in Blood and UrineArchives of Environmental Health: An International Journal, 1963