Head phantom for testing microwave systems for head imaging
- 1 December 2012
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21566097,p. 191-193
Abstract
The fabrication of a human head phantom for the use in microwave systems designed for head imaging is presented. Since most of the ongoing research activities on microwave imaging of human head use the band from 1 GHz to 4 GHz, the fabricated phantom is designed to emulate the electrical properties of human brain tissues across that band. A material that has electrical properties equal to those of the soft human brain tissues is manufactured using simple substances. A fabricated material that resembles the shape of human skull is used to contain the developed soft tissues. The measured properties of the fabricated phantom using a suitable dielectric probe are in good agreement with the properties of the real head tissues.Keywords
This publication has 19 references indexed in Scilit:
- COMPACT WIDEBAND ANTENNA FOR MICROWAVE IMAGING OF BRAINProgress in Electromagnetics Research C, 2012
- Validation of microwave radiometry for measuring the internal temperature profile of human tissueMeasurement Science and Technology, 2011
- Microwave Tomography for Brain Imaging: Feasibility Assessment for Stroke DetectionInternational Journal of Antennas and Propagation, 2008
- Directive Antenna for Ultrawideband Medical Imaging SystemsInternational Journal of Antennas and Propagation, 2008
- Detection of Hemorrhagic Hypointense Foci in the Brain on Susceptibility-Weighted ImagingAcademic Radiology, 2007
- Guidelines for the Early Management of Adults With Ischemic StrokeCirculation, 2007
- Imaging of acute strokeThe Lancet Neurology, 2006
- Compact ultra‐wideband planar tapered slot antenna for use in a microwave imaging systemMicrowave and Optical Technology Letters, 2006
- Design of a human-head-equivalent phantom for ISM 2.4-GHz applicationsMicrowave and Optical Technology Letters, 2005
- Medical imaging with a microwave tomographic scannerIEEE Transactions on Biomedical Engineering, 1990