Abstract
This paper establishes a standard method for measuring human tremor. The electronic instrument described is an application of this method. It solves the need for an effective and simple tremor-measuring instrument fit for wide distribution. This instrument consists of a piezoelectric accelerometer connected to an electronic circuit and to an LCD display. The signal is also analysed by a computer after accelerometer analogic/digital conversion in order to test the method. The tremor of 1079 healthy subjects was studied. Spectral analysis showed frequency peaks between 5.85 and 8.80 Hz. Chronic cigarette-smoking and coffee drinking did not modify the tremor as compared with controls. Relaxation session decreased tremor significantly in healthy subjects (P less than 0.01). This new tremor-measuring method opens new horizons in the understanding of physiological and pathological tremor, stress, anxiety and in the means to avoid or compensate them.

This publication has 20 references indexed in Scilit: