Radiosurgical Treatment Verification Using Removable Megavoltage Radiation Detectors
Open Access
- 29 November 2017
- journal article
- research article
- Published by Springer Nature in Cureus
- Vol. 9 (11) , e1889
Abstract
Accurate dose delivery is critical to the success of stereotactic radiosurgery. Unfortunately, verification of the accuracy of treatment delivery remains a challenging problem. Existing radiosurgery delivery paradigms are limited in their ability to verify the accurate delivery of radiation beams using data sampled from the beam after it has traversed the patient. The Zap-X Radiosurgery System (Zap Surgical Systems, San Carlos, CA) addresses this issue by implementing a fully integrated treatment delivery system that utilizes a factory commissioned megavoltage (MV) imager to measure the transmitted beam. The measured intensity is then compared with an expected value in order to confirm that treatment is proceeding as expected. The purpose of this study was to evaluate a prototype system and investigate the accuracy of an attenuation model used in generating the expected transmitted intensity values. A prototype MV imager was used to measure transmitted beam intensities at various exposure levels and through several thicknesses of solid water. The data were used to evaluate imager linearity and model accuracy. Experimental results indicate that a quadratic attenuation model is appropriate for predicting beam attenuation and that the imager exhibits excellent dose linearity. The MV imager system is shown to be capable of accurately acquiring the data needed to confirm treatment validity.Keywords
This publication has 9 references indexed in Scilit:
- Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variationsMedical Physics, 2015
- A Review of Stereotactic Radiosurgery Practice in the Management of Skull Base MeningiomasJournal of Neurological Surgery Part B: Skull Base, 2014
- Quality and safety considerations in stereotactic radiosurgery and stereotactic body radiation therapy: Executive summaryPractical Radiation Oncology, 2012
- Report of AAPM TG 135: Quality assurance for robotic radiosurgeryMedical Physics, 2011
- The role of stereotactic radiosurgery in the management of patients with newly diagnosed brain metastases: a systematic review and evidence-based clinical practice guidelineJournal of Neuro-Oncology, 2009
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Stereotactic radiosurgical treatment of brain metastasesJournal of Neurosurgery, 1992
- A 4-MV CT scanner for radiation therapy; spectral properties of the therapy beamMedical Physics, 1983