Photon-Counting Computed Tomography for Coronary Stent Imaging
- 9 April 2021
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 56 (10) , 653-660
Abstract
Objectives: The aim of this study was to assess in-stent lumen visibility and quantitative image characteristics of different coronary stents using a novel photon-counting detector (PCD) computed tomography (CT) system in comparison to a state-of-the-art energy-integrating detector (EID) CT scanner. Materials and Methods: In this in vitro phantom study, 28 different coronary stents ranging from 2.25 to 4.5 mm lumen diameter were expanded into plastic tubes filled with contrast agent. Stent-containing plastic tubes were positioned in a custom-made emulsion-filled phantom, which was inserted into an anthropomorphic phantom simulating a medium-sized patient. Computed tomography scans were acquired parallel to the scanners’ z axis using a novel cadmium telluride–based PCD CT system (SOMATOM CountPlus; Siemens Healthcare GmbH, Forchheim Germany), operating in 2 different modes (standard-resolution mode [SR] and ultra-high-resolution [UHR] mode), and a latest generation dual-source EID CT system (SOMATOM Force; Siemens Healthcare GmbH, Forchheim). CTDIvol-matched images were reconstructed with comparable convolution kernels and using the same reconstruction parameters. In-stent lumen visibility (in %), increase in in-stent attenuation (expressed as Δ in-stent CT attenuation), and image noise (in Hounsfield unit) were manually measured. Parts of the image analysis (in-stent lumen visibility) were additionally performed in an automated way. Differences were tested using Wilcoxon signed rank test. Results: The best in-stent lumen visibility was achieved with the PCD-UHR mode and the lowest noise levels with the PCD-SR mode. The median in-stent lumen visibility was significantly higher (P < 0.001) with PCD (SR, 66.7%; interquartile range [IQR], 63.3–72.3; UHR, 68.9%; IQR, 64.4–74.4) compared with EID (65.4%; IQR, 62.2–70.4). The Δ in-stent CT attenuation was significantly lower for PCD in both SR (78 HU; IQR, 46–108; P = 0.024) and UHR (85 HU; IQR, 59–113; P = 0.006) compared with EID (108 HU; IQR, 85–126). Image noise was significantly lower (P < 0.001) for PCD-SR (21 HU; IQR, 21–21) compared with EID images (25 HU; IQR, 24–25.0). Conclusions: The PCD provides superior in-stent lumen visibility and quantitative image characteristics when compared with conventional EID.Keywords
This publication has 33 references indexed in Scilit:
- Next generation coronary CT angiography: in vitro evaluation of 27 coronary stentsEuropean Radiology, 2014
- Multislice Computed Tomography Angiography for Noninvasive Assessment of the 18-Month Performance of a Novel Radiolucent Bioresorbable Vascular Scaffolding DeviceJACC, 2013
- Impact of Percutaneous Coronary Intervention Performance Reporting on Cardiac Resuscitation CentersCirculation, 2013
- A Contemporary View of Diagnostic Cardiac Catheterization and Percutaneous Coronary Intervention in the United StatesJACC, 2012
- Myocardial Infarction Due to Percutaneous Coronary InterventionNew England Journal of Medicine, 2011
- In-Stent Restenosis in the Drug-Eluting Stent EraJACC, 2010
- Image quality optimization and evaluation of linearly mixed images in dual‐source, dual‐energy CTMedical Physics, 2009
- A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trialThe Lancet, 2008
- Diagnostic accuracy of in-stent coronary restenosis detection with multislice spiral computed tomography: a meta-analysisEuropean Radiology, 2007
- Comparison of Coronary-Artery Bypass Surgery and Stenting for the Treatment of Multivessel DiseaseNew England Journal of Medicine, 2001