Towards a clinically applicable robotic assistance system for retinal vein cannulation
- 1 June 2016
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 284-291
Abstract
Retinal vein cannulation is a promising yet challenging eye surgical procedure currently not practised in clinic. Its goal is to inject a clot-dissolving drug into an occluded retinal vein. Given the scale and the fragility of retinal veins on one side and surgeons' limited positioning precision and force perception on the other side, performing this procedure manually is considered to be too risky. We previously developed a robotic system and a force-sensitive needle to assist surgeons in performing this procedure in a safe manner. In this paper we report on further technological developments and in-lab experiments on the path towards a clinically applicable system. The robotic system is extended with an alignment stage to pre-operatively position it with respect to the patient. It is demonstrated that an alignment accuracy of 0.4 mm can be obtained in less than 20 s. Further, a puncture detection algorithm is proposed for the force-sensitive needle. A 98% detection success rate is shown when cannulating retinal vessels of dissected porcine eyes. Finally, a junior retinal surgeon was invited to perform retinal vein cannulation with the aid of the developed technology on enucleated porcine eyes. The surgeon was able to cannulate and inject a fluid in 20/20 of the times when using the robot and the force-sensitive tool.Keywords
This publication has 21 references indexed in Scilit:
- Development and experimental validation of a force sensing needle for robotically assisted retinal vein cannulations2015
- Design of 4-DOF Parallelogram-Based RCM Mechanisms with a Translational DOF Implemented Distal from the End-Effector2014
- The introduction of a new robot for assistance in ophthalmic surgery2013
- Impact of Robotic Assistance on Precision of Vitreoretinal Surgical ProceduresPLOS ONE, 2013
- Characterization of Puncture Forces for Retinal Vein CannulationJournal of Medical Devices, 2011
- The Prevalence of Retinal Vein Occlusion: Pooled Data from Population Studies from the United States, Europe, Asia, and AustraliaOphthalmology, 2010
- Retinal Endovascular Lysis in Ischemic Central Retinal Vein OcclusionOphthalmology, 2007
- Retinal vein cannulation with prolonged infusion of tissue plasminogen activator (t-PA) for the treatment of experimental retinal vein occlusion in dogsAmerican Journal of Ophthalmology, 2004
- Dimensions of the human eye relevant to radiation protection (dosimetry)Physics in Medicine & Biology, 1975
- Diameters of the Retinal Vessels in Diabetic and Normal SubjectsDiabetes, 1969