Factors modulating 99mTc‐MAA planar lung dosimetry for 90Y radioembolization

Authors:

Abstract
Purpose: To investigate the accuracy and biases of predicted lung shunt fraction (LSF) and lung dose (LD) calculations via 99mTc‐macro‐aggregated albumin (99mTc‐MAA) planar imaging for treatment planning of 90Y‐microsphere radioembolization.Methods and materials: LSFs in 52 planning and LDs in 44 treatment procedures were retrospectively calculated, in consecutive radioembolization patients over a 2 year interval, using 99mTc‐MAA planar and SPECT/CT imaging. For each procedure, multiple planar LSFs and LDs were calculated using different: (1) contours, (2) views, (3) liver 99mTc‐MAA shine‐through compensations, and (4) lung mass estimations. The accuracy of each planar‐based LSF and LD methodology was determined by calculating the median (range) absolute difference from SPECT/CT‐based LSF and LD values, which have been demonstrated in phantom and patient studies to more accurately and reliably quantify the true LSF and LD values.Results: Standard‐of‐care LSF using geometric mean of lung and liver contours had median (range) absolute over‐estimation of 4.4 percentage points (pp) (0.9 to 11.9 pp) from SPECT/CT LSF. Using anterior views only decreased LSF errors (2.4 pp median, −1.1 to +5.7 pp range). Planar LD over‐estimations decreased when using single‐view versus geometric‐mean LSF (1.3 vs. 2.6 Gy median and 7.2 vs. 18.5 Gy maximum using 1000 g lung mass) but increased when using patient‐specific versus standard‐man lung mass (2.4 vs. 1.3 Gy median and 11.8 vs. 7.2 Gy maximum using single‐view LSF).Conclusions: Calculating planar LSF from lung and liver contours of a single view and planar LD using that same LSF and 1000 g lung mass was found to improve accuracy and minimize bias in planar lung dosimetry.

This publication has 19 references indexed in Scilit: