Biomechanical In Vitro Testing of Human Osteoporotic Lumbar Vertebrae Following Prophylactic Kyphoplasty With Different Candidate Materials
Authors:
- Robert PflugmacherRPRobert Pflugmacher
- Frank KandzioraFKFrank Kandziora
- Andreas EwertAEAndreas Ewert
- Georg DudaGDGeorg Duda
- Thomas MittlmeierTMThomas Mittlmeier
- 1 June 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 32 (13) , 1400-1405
Abstract
Biomechanical study of human osteoporotic lumbar vertebrae following prophylactic kyphoplasty. To evaluate the potential benefits of different resorbable candidate materials for use in prophylactic kyphoplasty compared with the behavior of polymethylmethacrylate cement. Kyphoplasty using PMMA bone cement for the stabilization of fractured osteoporotic vertebrae has been established as a useful clinical tool. In several studies, consecutive compression fractures have been reported in vertebrae caudal or cranial to those augmented with bone cement. Consequently, some physicians have begun to treat adjacent vertebrae by means of prophylactic augmentation. Biomechanical in vitro testing was performed on 40 human osteoporotic nonfractured lumbar vertebrae. Three types of bone cement (PMMA, 2 different calcium phosphate cements) and one silicon derivative were assessed during compressive and cyclic sinusoidal testing. Each candidate material was applied bipedicularly under fluoroscopic control. Differing processing qualities of the materials led to substantial differences during cement injection, in particular in the amount of cement filling of the vertebrae. However, in comparison to native vertebrae, augmented specimens showed significantly higher compressive failure. No significant differences between vertebral bodies treated with PMMA and those treated with either type of calcium phosphate cement were documented. The biomechanical properties of the vertebrae could not be significantly improved by the silicon derivative. This study demonstrated that calcium phosphate cements displayed identical behavior to PMMA cement with respect to in vitro mechanical qualities. Consequently, from a mechanical viewpoint, calcium phosphate cements may be used in addition to PMMA cement for kyphoplasty and prophylactic kyphoplasty. Silicon derivatives are apparently not recommendable as candidate materials for kyphoplasty.Keywords
This publication has 14 references indexed in Scilit:
- Calcium phosphate cement–based vertebroplasty compared with conservative treatment for osteoporotic compression fractures: a matched case–control studyJournal of Neurosurgery: Spine, 2006
- The Risk of New Osteoporotic Vertebral Compression Fractures in the Year after Percutaneous VertebroplastyJournal of Vascular and Interventional Radiology, 2006
- Radiological Follow-up of New Compression Fractures Following Percutaneous VertebroplastyCardioVascular and Interventional Radiology, 2005
- Preliminary outcomes and efficacy of the first 360 consecutive kyphoplasties for the treatment of painful osteoporotic vertebral compression fracturesThe Spine Journal, 2005
- CT volumetry of intravertebral cement after kyphoplasty. Comparison of polymethylmethacrylate and calcium phosphate in a 12-month follow-upEuropean Radiology, 2005
- Vertebroplastik bei hochgradiger OsteoporoseDer Orthopäde, 2004
- Adjacent vertebral failure after vertebroplastyThe Journal of Bone and Joint Surgery. British volume, 2002
- Augmentation of mechanical properties in osteoporotic vertebral bones – a biomechanical investigation of vertebroplasty efficacy with different bone cementsEuropean Spine Journal, 2001
- Long‐term observations of vertebral osteoporotic fractures treated by percutaneous vertebroplastyRheumatology, 2000
- The diagnosis of osteoporosisJournal of Bone and Mineral Research, 1994