J‐difference lactate editing at 3.0 Tesla in the presence of strong lipids
Authors:
Open Access
- 24 November 2008
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 28 (6) , 1492-1498
Abstract
Purpose To implement in vivo detection of lactate in the presence of lipids by proton magnetic resonance spectroscopy at a 3 Tesla (T) field strength for potential applications in human tumors outside of the brain. Materials and Methods The BASING J‐difference sequence was implemented in the presence of high lipid concentrations in phantoms and in vivo at 3 Tesla. Results The effectiveness of the lactate editing scheme is demonstrated in phantoms containing both lactate and lipids and in vivo in ischemic induced human muscle. Conclusion The ability of the BASING J‐difference technique to detect lactate in the presence of strong lipid signals outside the brain at 3T is feasible. This robust technique should permit noninvasive lactate measurements in human tumors to investigate its potential as a prognostic indicator. J. Magn. Reson. Imaging 2008;28:1492–1498.Keywords
This publication has 30 references indexed in Scilit:
- Brain Metabolite Levels Assessed by Lactate-Edited MR Spectroscopy in Premature Neonates with and without Pentobarbital SedationAmerican Journal of Neuroradiology, 2008
- Direct noninvasive quantification of lactate and high energy phosphates simultaneously in exercising human skeletal muscle by localized magnetic resonance spectroscopyMagnetic Resonance in Medicine, 2007
- Optimized detection of lactate at high fields using inner volume saturationMagnetic Resonance in Medicine, 2006
- Metabolic classification of human rectal adenocarcinomas: a novel guideline for clinical oncologists?Journal of Cancer Research and Clinical Oncology, 2003
- Acute hypoxia increases the aggressive characteristics and survival properties of prostate cancer cellsThe Prostate, 2002
- Improved water and lipid suppression for 3D PRESS CSI using rf band selective inversion with gradient dephasing (basing)Magnetic Resonance in Medicine, 1997
- Practical implementation of single-voxel double-quantum editing on a whole-body NMR spectrometer: Localized monitoring of lactate in the human leg during and after exerciseMagnetic Resonance in Medicine, 1996
- Incorporating lactate/lipid discrimination into a spectroscopic imaging sequenceMagnetic Resonance in Medicine, 1993
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991
- The use of finite impulse response filters in pulse designMagnetic Resonance in Medicine, 1989