Urinary creatinine estimation as an important analytical limitation of semi-quantitative albumin-to-creatinine ratio strip testing: a diagnostic accuracy study
Authors:
Open Access
- 29 October 2026
- journal article
- research article
- Published by Frontiers Media SA in Frontiers in Medical Technology
- Vol. 8, 1913468
Abstract
Background: Albuminuria is a key biomarker for chronic kidney disease and cardiovascular risk. Point-of-care urine albumin-to-creatinine ratio (UACR) testing may improve access to chronic kidney disease screening. However, limited evidence is available regarding the agreement of the individual albumin and creatinine components of semi-quantitative UACR strip tests with quantitative laboratory measurements. Methods: We evaluated 400 spot urine specimens using the Dirui H13-Cr semi-quantitative UACR strip and compared results with quantitative immunoturbidimetric UACR measurement performed on a Roche Cobas c503 analyzer. Agreement was assessed using Cohen's kappa coefficients, including category-specific agreement according to the KDIGO A1, A2, A3 albuminuria categories, and diagnostic performance was evaluated using sensitivity, specificity, predictive values, likelihood ratios, and diagnostic odds ratio (DOR). Results: Agreement between strip-based and quantitative measurements was substantial for urinary albumin ( κ =0.691; 95% CI, 0.632–0.751) but considerably lower for urinary creatinine ( κ =0.377; 95% CI, 0.331–0.423). Despite this disparity, UACR classification demonstrated substantial agreement ( κ =0.627; 95% CI, 0.541–0.714). The findings suggest that semi-quantitative creatinine estimation is the primary analytical determinant of UACR misclassification. For detection of abnormal albuminuria (UACR ≥30 mg/g), the strip achieved a sensitivity of 83.66% and specificity of 83.22%. The PPV was 89.96% and NPV was 73.91%. Category-specific analysis showed that 131 of 177 A2 specimens (74.0%) were classified as A2 by the strip, whereas 64 of 80 A3 specimens (80.0%) were classified as A2 rather than A3. Conclusion: The Dirui H13-Cr strip demonstrated acceptable performance for albuminuria screening. However, semi-quantitative urinary creatinine estimation represented an important analytical limitation affecting UACR classification accuracy. Future development of strip-based UACR technologies should prioritize improved creatinine measurement to enhance diagnostic performance.Keywords
This publication has 30 references indexed in Scilit:
- Lifestyle and Metabolic Health, Low-Grade to Clinically Elevated Albuminuria, and Cardiovascular-Renal DiseasesJACC: Asia, 2026
- Development of a smartphone based spectrometer for high-resolution urinalysisScientific Reports, 2026
- A passive flow microreactor for urine creatinine testMicrosystems & Nanoengineering, 2025
- KDIGO 2024 Clinical Practice Guideline for the Evaluation and Management of Chronic Kidney DiseaseKidney International, 2024
- Diagnostic accuracy of urine dipstick testing for albumin-to-creatinine ratio and albuminuria: A systematic review and meta-analysisHeliyon, 2021
- Evaluation the diagnostic accuracy of albuminuria detection in semi-quantitative urinalysisClinica Chimica Acta, 2020
- Prevalence of micro albuminuria and diagnostic accuracy of urine dipstick for the screening of diabetic nephropathy in type 2 diabetes patientsBiocatalysis and Agricultural Biotechnology, 2019
- Evaluation of the URiSCAN Super Cassette ACR Semiquantitative Urine Dipstick for Microalbuminuria ScreeningJournal of Clinical Laboratory Analysis, 2014
- Association of estimated glomerular filtration rate and albuminuria with all-cause and cardiovascular mortality in general population cohorts: a collaborative meta-analysisThe Lancet, 2010
- Diagnostic accuracy of the urinary albumin: creatinine ratio determined by the CLINITEK Microalbumin and DCA 2000+ for the rule-out of albuminuria in chronic kidney diseaseClinica Chimica Acta, 2009