Determination of aluminum in moisturizing body lotions using graphite furnace atomic absorption spectrometry
- 28 October 2015
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Analytical Methods
- Vol. 7 (22) , 9636-9640
Abstract
A method has been developed for the determination of Al in commercial moisturizing body lotions using high-resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GF AAS). The most sensitive absorption line at 309.271 nm has been used for all determinations. The samples were prepared via a suspension with 0.3 mol L−1 nitric acid. Comparisons were made with pyrolysis and atomization curves without a modifier and using ruthenium and zirconium as permanent modifiers, to reduce the interaction of Al with the graphite surface and to improve the atomization efficiency of Al. 400 μg Zr was finally chosen as the best permanent modifier. The limits of detection and quantification were 30 ng g−1 and 95 ng g−1 Al, respectively. The proposed method was applied for 15 commercial moisturizing body lotions with results ranging between 0.17 ± 0.05 μg g−1 and 11.8 ± 0.43 μg g−1. Analyte addition tests have been used to evaluate the accuracy of the method and the recovery ranged between 103 and 110%, indicating that there was no serious matrix effect.Keywords
Funding Information
- Conselho Nacional de Desenvolvimento Científico e Tecnológico (305085/2011-5)
This publication has 16 references indexed in Scilit:
- Selective and sensitive spectrophotometric method for determination of sub-micro-molar amounts of aluminium ionJournal of Hazardous Materials, 2008
- Feasibility of peak volume, side pixel and multiple peak registration in high-resolution continuum source atomic absorption spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 2007
- Current Status of Direct Solid Sampling for Electrothermal Atomic Absorption Spectrometry—A Critical Review of the Development between 1995 and 2005Applied Spectroscopy Reviews, 2006
- Aluminum: new recognition of an old problemCurrent Opinion in Pharmacology, 2005
- Development of methodologies to determine aluminum, cadmium, chromium and lead in drinking water by ET AAS using permanent modifiersTalanta, 2004
- Validação em métodos cromatográficos e eletroforéticosQuímica Nova, 2004
- Study of parameters associated with the determination of cadmium in lake sediments by slurry sampling electrothermal atomic absorption spectrometryAnalytica Chimica Acta, 2004
- Slurry sampling for direct analysis of solid materials by electrothermal atomic absorption spectrometry (ETAAS). A literature review from 1990 to 2000Talanta, 2002
- Direct analysis of solid samples by GFAAS - determination of trace heavy metals in barytesAnalytical and Bioanalytical Chemistry, 1999