Interleaved narrow‐band PRESS sequence with adiabatic spatial‐spectral refocusing pulses for 1H MRSI at 7T
Open Access
- 21 April 2008
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 59 (5) , 973-979
Abstract
Proton magnetic resonance spectroscopic imaging (1H MRSI) is a useful technique for measuring metabolite levels in vivo, with Choline (Cho), Creatine (Cre), and N‐Acetyl‐Aspartate (NAA) being the most prominent MRS‐detectable brain biochemicals. 1H MRSI at very high fields, such as 7T, offers the advantages of higher SNR and improved spectral resolution. However, major technical challenges associated with high‐field systems, such as increased B 1 and B 0 inhomogeneity as well as chemical shift localization (CSL) error, degrade the performance of conventional 1H MRSI sequences. To address these problems, we have developed a Position Resolved Spectroscopy (PRESS) sequence with adiabatic spatial‐spectral (SPSP) refocusing pulses, to acquire multiple narrow spectral bands in an interleaved fashion. The adiabatic SPSP pulses provide magnetization profiles that are largely invariant over the 40% B 1 variation measured across the brain at 7T. Additionally, there is negligible CSL error since the transmit frequency is separately adjusted for each spectral band. in vivo 1H MRSI data were obtained from the brain of a normal volunteer using a standard PRESS sequence and the interleaved narrow‐band PRESS sequence with adiabatic refocusing pulses. In comparison with conventional PRESS, this new approach generated high‐quality spectra from an appreciably larger region of interest and achieved higher overall SNR. Magn Reson Med 59:973–979, 2008.Keywords
This publication has 23 references indexed in Scilit:
- Fast 1 H spectroscopic imaging using steady state free precession and spectral–spatial RF pulsesMagnetic Resonance Materials in Physics, Biology and Medicine, 2006
- In Vivo Magnetic Resonance Spectroscopy in CancerAnnual Review of Biomedical Engineering, 2005
- Sequence design for magnetic resonance spectroscopic imaging of prostate cancer at 3 TMagnetic Resonance in Medicine, 2005
- Design of symmetric‐sweep spectral‐spatial RF pulses for spectral editingMagnetic Resonance in Medicine, 2004
- Single‐shot two‐echo technique for simultaneous measurement of GABA and creatine in the human brain in vivoMagnetic Resonance in Medicine, 2004
- Improved solvent suppression and increased spatial excitation bandwidths for three-dimensional press CSI using phase-compensating spectral/spatial spin-echo pulsesJournal of Magnetic Resonance Imaging, 1997
- Imaging of the active B1 field in vivoMagnetic Resonance in Medicine, 1996
- Two‐dimensional selective adiabatic pulsesMagnetic Resonance in Medicine, 1992
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991
- Spatial Localization in NMR Spectroscopy in VivoAnnals of the New York Academy of Sciences, 1987