Hip screw migration testing: First results for hip screws and helical blades utilizing a new oscillating test method
Open Access
- 9 September 2010
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 29 (5) , 760-766
Abstract
Despite continued improvement in the methods and devices used to treat intertrochanteric fractures, there remains an unacceptable amount of failures. The cut‐out rate for hip screws has been recorded up to 8.3%. To evaluate the migration of different implants under physiological loads, a multiplanar biomechanical test method for hip screws was developed, the first to incorporate a simulation of the human gait cycle by an oscillating flexion/extension movement of the test device. The new method was used to compare different hip screw and blade designs with respect to their directional migration resistance. The test method generated failure modes that were consistent with those observed clinically. Under cyclic loading, the hip screws migrated predominantly in a cephalad direction. In contrast, the helical blades exhibited a distinct migration in their axial direction. The Gamma3 hip screw design showed a significantly higher migration resistance compared with other screw and helical blade designs. The results demonstrate the ability of hip screws to significantly reduce axial migration and prevent cut‐out under simulated walking loads. Further, the new multiplanar test method creates a physiological environment that can be used to optimize designs for intertrochanteric fracture fixation. © 2010 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 29:760–766, 2011Keywords
This publication has 23 references indexed in Scilit:
- Salvage of Failed Hip Fracture FixationJournal of Orthopaedic Trauma, 2009
- Comparison of migration behavior between single and dual lag screw implants for intertrochanteric fracture fixationJournal of Orthopaedic Surgery and Research, 2009
- Compressive properties of commercially available polyurethane foams as mechanical models for osteoporotic human cancellous boneBMC Musculoskeletal Disorders, 2008
- Lag screws for hip fracture fixation: Evaluation of migration resistance under simulated walkingJournal of Orthopaedic Research, 2005
- Osteoporotic Pertrochanteric Hip FracturesJournal of Bone and Joint Surgery, 2004
- Implant-related complications in the treatment of unstable intertrochanteric fractures: meta-analysis of dynamic screw-plate versus dynamic screw-intramedullary nail devicesInternational Orthopaedics, 2003
- Failure of intertrochanteric fracture fixation with a dynamic hip screw in relation to pre-operative fracture stability and osteoporosisInternational Orthopaedics, 2001
- Immediate Weight-Bearing After Treatment of a Comminuted Fracture of the Femoral Shaft with a Statically Locked Intramedullary Nail*Journal of Bone and Joint Surgery, 1999
- The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip.Journal of Bone and Joint Surgery, 1995