Traditional methods for measuring in-pile temperatures degrade above 1100°C. Hence, the Idaho National Laboratory (INL) initiated a project to explore the use of specialized thermocouples for high temperature in-pile applications. Efforts to develop, fabricate, and evaluate specialized high-temperature thermocouples for in-pile applications suggest that several material combinations are viable. Tests show that several low-neutron cross-section candidate materials resist material interactions and remain ductile at high temperatures. In addition, results indicate that the candidate thermoelements have a thermoelectric response that is single-valued and repeatable with acceptable resolution. The selection of the thermocouple materials depends on desired peak temperature and accuracy requirements. For applications at or above 1600°C, tests indicate that thermocouples having doped molybdenum and Nb-1%Zr thermoelement wires, HfO2 insulation, and a Nb-1%Zr sheath could be used.INL has worked to optimize this the...Keywords