Using combined x-ray and MR imaging for prostate I-125 post-implant dosimetry: phantom validation and preliminary patient work
Authors:
- M E MiquelMMM E Miquel
- K S RhodeK S Rhode
- P L AcherPAP L Acher
- N D MacDougallNMN D MacDougall
- J BlackallJBJ Blackall
- R P GastonRGR P Gaston
- S HegdeSHS Hegde
- S L MorrisSMS L Morris
- R BeaneyRBR Beaney
- C DeehanCDC Deehan
- R PopertRPR Popert
- S F KeevilSKS F Keevil
- 8 February 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (5) , 1129-1137
Abstract
Post-implantation dosimetry is an important element of permanent prostate brachytherapy. This process relies on accurate localization of implanted seeds relative to the surrounding organs. Localization is commonly achieved using CT images, which provide suboptimal prostate delineation. On MR images, conversely, prostate visualization is excellent but seed localization is imprecise due to distortion and susceptibility artefacts. This paper presents a method based on fused MR and x-ray images acquired consecutively in a combined x-ray and MRI interventional suite. The method does not rely on any explicit registration step but on a combination of system calibration and tracking. A purpose-built phantom was imaged using MRI and x-rays, and the images were successfully registered. The same protocol was applied to three patients where combining soft tissue information from MRI with stereoscopic seed identification from x-ray imaging facilitated post-implant dosimetry. This technique has the potential to improve on dosimetry using either CT or MR alone.Keywords
This publication has 23 references indexed in Scilit:
- European Collaborative Group on Prostate Brachytherapy: Preliminary Report in 1175 PatientsEuropean Urology, 2004
- MRI-guided HDR prostate brachytherapy in standard 1.5T scannerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Visualization and tracking of an inflatable balloon catheter using SSFP in a flow phantom and in the heart and great vessels of patientsMagnetic Resonance in Medicine, 2004
- Prostate BrachytherapyCancer and Metastasis Reviews, 2002
- A software system for interventional magnetic resonance image-guided prostate brachytherapyComputer Aided Surgery, 2000
- Prostate seed implant quality assessment using MR and CT image fusionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1999
- Prostate brachytherapyProstate Cancer and Prostatic Diseases, 1998
- Source localization following permanent transperineal prostate interstitial brachytherapy using magnetic resonance imagingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1997
- Evaluation of permanent I-125 prostate implants using radiography and magnetic resonance imagingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1997
- Effect of fiducial marker localization on stereotactic target coordinate calculation in CT slices and radiographsPhysics in Medicine & Biology, 1994