Ventilator-Related Adverse Events: A Taxonomy and Findings From 3 Incident Reporting Systems
Authors:
Open Access
- 26 January 2016
- journal article
- Published by SAGE Publications in Respiratory Care
- Vol. 61 (5) , 621-631
Abstract
BACKGROUND: In 2009, researchers from Johns Hopkins University's Armstrong Institute for Patient Safety and Quality; public agencies, including the FDA; and private partners, including the Emergency Care Research Institute and the University HealthSystem Consortium (UHC) Safety Intelligence Patient Safety Organization, sought to form a public-private partnership for the promotion of patient safety (P5S) to advance patient safety through voluntary partnerships. The study objective was to test the concept of the P5S to advance our understanding of safety issues related to ventilator events, to develop a common classification system for categorizing adverse events related to mechanical ventilators, and to perform a comparison of adverse events across different adverse event reporting systems. METHODS: We performed a cross-sectional analysis of ventilator-related adverse events reported in 2012 from the following incident reporting systems: the Pennsylvania Patient Safety Authority's Patient Safety Reporting System, UHC's Safety Intelligence Patient Safety Organization database, and the FDA's Manufacturer and User Facility Device Experience database. Once each organization had its dataset of ventilator-related adverse events, reviewers read the narrative descriptions of each event and classified it according to the developed common taxonomy. RESULTS: A Pennsylvania Patient Safety Authority, FDA, and UHC search provided 252, 274, and 700 relevant reports, respectively. The 3 event types most commonly reported to the UHC and the Pennsylvania Patient Safety Authority's Patient Safety Reporting System databases were airway/breathing circuit issue, human factor issues, and ventilator malfunction events. The top 3 event types reported to the FDA were ventilator malfunction, power source issue, and alarm failure. CONCLUSIONS: Overall, we found that (1) through the development of a common taxonomy, adverse events from 3 reporting systems can be evaluated, (2) the types of events reported in each database were related to the purpose of the database and the source of the reports, resulting in significant differences in reported event categories across the 3 systems, and (3) a public-private collaboration for investigating ventilator-related adverse events under the P5S model is feasible.Keywords
This publication has 25 references indexed in Scilit:
- Impacto de una intervención de mejora de calidad sobre la incidencia de extubaciones no planeadas en una unidad de cuidados intensivos pediátricosArchivos Argentinos de Pediatria, 2013
- Unplanned Extubation in the Neonatal ICU: A Systematic Review, Critical Appraisal, and Evidence-Based RecommendationsRespiratory Care, 2013
- Patient-ventilator dyssynchrony during assisted invasive mechanical ventilation.2012
- An Interdisciplinary Initiative to Reduce Unplanned Extubations in Pediatric Critical Care UnitsPediatrics, 2012
- Not-so-Trivial Pursuit: Mechanical Ventilation Risk ReductionAmerican Journal of Critical Care, 2009
- Risk factors for delirium in intensive care patients: a prospective cohort studyCritical Care, 2009
- Reducing Health Care Hazards: Lessons From The Commercial Aviation Safety TeamHealth Affairs, 2009
- Unplanned extubation in a paediatric intensive care unit: impact of a quality improvement programmeAnaesthesia, 2008
- Characteristics Associated With Unplanned Extubations in a Surgical Intensive Care UnitAmerican Journal of Critical Care, 2008
- Unplanned extubation in adult critical care. Quality improvement and education payoff.2004