Effect of firing rate on the performance of shock wave lithotriptors
Open Access
- 21 November 2008
- journal article
- research article
- Published by Wiley in BJU International
- Vol. 102 (11) , 1681-1686
Abstract
OBJECTIVE: To determine the mechanism that underlies the effect of shock wave (SW) rate on the performance of clinical lithotripters. MATERIALS AND METHODS: The effect of firing rate on the pressure characteristics of SWs was assessed using a fibre‐optic probe hydrophone (FOPH 500, RP Acoustics, Leutenbach, Germany). Shock waves were fired at slow (5–27 SW/min) and fast (100–120 SW/min) rates using a conventional high‐pressure lithotriptor (DoLi‐50, Dornier MedTech America, Inc., Kennesaw, GA, USA), and a new low‐pressure lithotriptor (XX‐ES, Xi Xin Medical Instruments Co. Ltd, Suzhou, PRC). A digital camcorder (HDR‐HC3, Sony, Japan) was used to record cavitation fields, and an ultrafast multiframe high‐speed camera (Imacon 200, DRS Data & Imaging Systems, Inc., Oakland, NJ, USA) was used to follow the evolution of bubbles throughout the cavitation cycle. RESULTS: Firing rate had little effect on the leading positive‐pressure phase of the SWs with the DoLi lithotriptor. A slight reduction (≈7%) of peak positive pressure (P+) was detected only in the very dense cavitation fields (≈1000 bubbles/cm 3 ) generated at the fastest firing rate (120 SW/min) in nondegassed water. The negative pressure of the SWs, on the other hand, was dramatically affected by firing rate. At 120 SW/min the peak negative pressure was reduced by ≈84%, the duration and area of the negative pressure component was reduced by ≈80% and ≈98%, respectively, and the energy density of negative pressure was reduced by >99%. Whereas cavitation bubbles proliferated at fast firing rates, HS‐camera images showed the bubbles that persisted between SWs were very small (<10 µm). Similar results were obtained with the XX‐ES lithotriptor but only after recognizing a rate‐dependent charging artefact with that machine. CONCLUSION: Increasing the firing rate of a lithotriptor can dramatically reduce the negative pressure component of the SWs, while the positive pressure remains virtually unaffected. Cavitation increases as the firing rate is increased but as the bubbles collapse, they break into numerous microbubbles that, because of their very small size, do not pose a barrier to the leading positive pressure of the next SW. These findings begin to explain why stone breakage in SWL becomes less efficient as the firing rate is increased.Keywords
This publication has 28 references indexed in Scilit:
- The Acute and Long-Term Adverse Effects of Shock Wave LithotripsySeminars in Nephrology, 2008
- Renal injury during shock wave lithotripsy is significantly reduced by slowing the rate of shock wave deliveryBJU International, 2007
- Dual‐head lithotripsy in synchronous mode: acute effect on renal function and morphology in the pigBJU International, 2007
- Air Pockets Trapped During Routine Coupling in Dry Head Lithotripsy Can Significantly Decrease the Delivery of Shock Wave EnergyJournal of Urology, 2006
- Detection of Significant Variation in Acoustic Output of an Electromagnetic LithotriptorJournal of Urology, 2006
- Why Stones Break Better at Slow Shockwave Rates Than at Fast Rates: In Vitro Study with a Research Electrohydraulic LithotripterJournal of Endourology, 2006
- SLOW VERSUS FAST SHOCK WAVE LITHOTRIPSY RATE FOR UROLITHIASIS: A PROSPECTIVE RANDOMIZED STUDYJournal of Urology, 2005
- Stone Fragmentation During Shock Wave Lithotripsy is Improved by Slowing the Shock Wave Rate: Studies With a New Animal ModelJournal of Urology, 2002
- Shockwave Frequency Affects Fragmentation in a Kidney Stone ModelJournal of Endourology, 2000
- Implications of cavitation phenomena for shot intervals in extracorporeal shock wave lithotripsyBritish Journal of Urology, 1995