Accuracy of Accusport for measurement of lactate concentrations in equine blood and plasma

Authors:

Abstract
Summary: The aim of this study was to investigate correlations between lactate concentrations in equine whole blood and plasma measured with Accusport 1 and Yellow Springs Instruments (YSI 2 ) (2300) methods. The effect of packed cell volume (PCV) on the accuracy of Accusport was also investigated. Blood samples were collected from Thoroughbred horses at 5–10 min intervals after a treadmill exercise test. Blood was added to NaEDTA (for PCV measurement) and to 2 tubes containing lithium heparin anticoagulant (for lactate assays). At concentrations greater than 10 mmol/1, Accusport 1 greatly underestimated the blood lactate concentration. Accusport 1 also underestimated blood lactate concentration when PCV were greater than 53%. Accusport 1 accurately measured plasma lactate concentration in the range 0.8–20 mmol/1.