Challenges in modeling the Agility multileaf collimator in treatment planning systems and current needs for improvement
Authors:
Open Access
- 27 October 2022
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 49 (12) , 7404-7416
Abstract
Background: The Agility multileaf collimator (MLC) mounted in Elekta linear accelerators features some unique design characteristics, such as large leaf thickness, eccentric curvature at the leaf tip, and defocused leaf sides (‘tilting'). These characteristics offer several advantages but modeling them in treatment planning systems (TPSs) is challenging.Purpose: The goals of this study were to investigate the challenges faced when modeling the Agility in two commercial TPSs (Monaco and RayStation) and to explore how the implemented MLC models could be improved in the future.Methods: Four linear accelerators equipped with the Agility, located at different centers, were used for the study. Three centers use the RayStation TPS and the other one uses Monaco. For comparison purposes, data from four Varian linear accelerators with the Millennium 120 MLC were also included. Average doses measured with asynchronous sweeping gap tests were used to characterize and compare the characteristics of the Millennium and the Agility MLCs and to assess the MLC model in the TPSs. The FOURL test included in the ExpressQA package, provided by Elekta, was also used to evaluate the tongue‐and‐groove with radiochromic films. Finally, raytracing was used to investigate the impact of the MLC geometry and to understand the results obtained for each MLC.Results: The geometry of the Agility produces dosimetric effects associated with the rounded leaf end up to a distance 20 mm away from the leaf tip end measured at the isocenter plane. This affects the tongue‐and‐groove shadowing, which progressively increases along the distance to the tip end. The RayStation and Monaco TPSs did not account for this effect, which made trade‐offs in the MLC parameters necessary and greatly varied the final MLC parameters used by different centers. Raytracing showed that these challenging leaf tip effects were directly related to the MLC geometry and that the characteristics mainly responsible for the large leaf tip effects of the Agility were its tilting design and its small source‐to‐collimator distance.Conclusions: The MLC models implemented in RayStation and Monaco could not accurately reproduce the leaf tip effects for the Agility. Therefore, trade‐offs are needed and the optimal MLC parameters are dependent on the specific characteristics of treatment plans. Refining the MLC models for the Agility to better approximate the measured leaf tip and tongue‐and‐groove effects would extend the validity of the MLC model, reduce the variability in the MLC parameters used by the community, and facilitate the standardization of the MLC configuration process.Keywords
Funding Information
- Department for Business, Energy and Industrial Strategy, UK Government
This publication has 35 references indexed in Scilit:
- Evaluating IMRT and VMAT dose accuracy: Practical examples of failure to detect systematic errors when applying a commonly used metric and action levelsMedical Physics, 2013
- An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scanMedical Physics, 2012
- Monte Carlo implementation, validation, and characterization of a 120 leaf MLCMedical Physics, 2011
- On the use of the MLC dosimetric leaf gap as a quality control tool for accurate dynamic IMRT deliveryMedical Physics, 2011
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- IMRT Delivery Performance With a Varian Multileaf CollimatorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- A virtual photon source model of an Elekta linear accelerator with integrated mini MLC for Monte Carlo based IMRT dose calculationPhysics in Medicine & Biology, 2007
- An experimental investigation into the radiation field offset of a dynamic multileaf collimatorPhysics in Medicine & Biology, 2006
- Verification of a rounded leaf-end MLC model used in a radiotherapy treatment planning systemPhysics in Medicine & Biology, 2006
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998