Demonstration of a memory calibration method in water isotope measurement by laser spectroscopy
- 4 December 2019
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 34 (8) , e8689
Abstract
Rationale: Measuring δ 18 O and δ 2 H values in water using wavelength‐scanned cavity ring down spectroscopy (WS‐CRDS) requires multiple injections of up to six (and sometimes eight or more) of one sample to remove the memory effect, which decreases the sample throughput and increases the consumables cost. Thus, improved methods for removing the memory effect are required. Methods: We calculated the memory coefficients by sequential WS‐CRDS measurement of two lab standard waters with isotopic differences, and used them to establish calibration equations. We then used these equations to correct the measured δ 18 O and δ 2 H values by removing the memory effect, instead of using multiple injections in the routine daily measurements. Results: By using this method, the number of injections per sample was reduced to one. The reproducibility (one standard deviation) of the δ 18 O and δ 2 H values obtained for quality control sample was less than 0.05‰ and 0.5‰ for an annual average, respectively. Conclusions: By measuring the memory coefficients and establishing the calibration equations, a highly effective method was developed for determining the δ 18 O and δ 2 H values of water, which could significantly improve sample throughput for liquid water dual isotope measurement without sacrificing the precision.Keywords
Funding Information
- National Natural Science Foundation of China (41771043, 41771085)
This publication has 15 references indexed in Scilit:
- Oxygen isotope measurements of seawater (18O/16O): A comparison of cavity ring‐down spectroscopy (CRDS) and isotope ratio mass spectrometry (IRMS)Limnology and Oceanography: Methods, 2015
- LIMS for Lasers 2015 for achieving long‐term accuracy and precision of δ 2 H, δ 17 O, and δ 18 O of waters using laser absorption spectrometryRapid Communications in Mass Spectrometry, 2015
- Stable Isotope Analysis of Saline Water Samples on a Cavity Ring-down Spectroscopy InstrumentEnvironmental Science & Technology, 2014
- Approaches for Achieving Long-Term Accuracy and Precision of δ18O and δ2H for Waters Analyzed using Laser Absorption SpectrometersEnvironmental Science & Technology, 2013
- Reducing and correcting for contamination of ecosystem water stable isotopes measured by isotope ratio infrared spectroscopyRapid Communications in Mass Spectrometry, 2011
- Analysis of the hydrogen and oxygen stable isotope ratios of beverage waters without prior water extraction using isotope ratio infrared spectroscopyRapid Communications in Mass Spectrometry, 2010
- Demonstration of high‐precision continuous measurements of water vapor isotopologues in laboratory and remote field deployments using wavelength‐scanned cavity ring‐down spectroscopy (WS‐CRDS) technologyRapid Communications in Mass Spectrometry, 2009
- A cavity ring-down analyzer for measuring atmospheric levels of methane, carbon dioxide, and water vaporApplied Physics B, 2008
- High-Precision Laser Spectroscopy D/H and 18O/16O Measurements of Microliter Natural Water SamplesAnalytical Chemistry, 2007
- An automated system for hydrogen isotope analysis of waterChemical Geology, 1998