Comparison of the St. Petersburg Phantom with a BOMAB Phantom in the ORTEC StandFast Whole Body Counter: A Monte Carlo Simulation
Authors:
- Gary H. KramerGKGary H. Kramer
- Kevin CapelloKCKevin Capello
- Jeremy SungJSJeremy Sung
- 1 May 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Health Physics
- Vol. 94 (5) , S78-S82
Abstract
Three sizes of the St. Petersburg phantom have been compared to six sizes of BOMAB phantoms measured by a virtual StandFast whole body counter using Monte Carlo simulations to investigate if the counting efficiencies are equivalent. This work shows that previously published data comparing the Reference Man sized phantom at 662 keV is supported; however, the simulations also show that the smaller sized St. Petersburg phantoms do not agree well with BOMAB phantoms. It is concluded that, compared with BOMAB phantoms, the St. Petersburg phantoms are system dependent and that they should be validated over a wide photon energy range against corresponding BOMAB phantoms prior to their use for calibrating whole body counters.Keywords
This publication has 6 references indexed in Scilit:
- THE STANDFAST WHOLE BODY COUNTER: EFFICIENCY AS A FUNCTION OF BOMAB PHANTOM SIZE AND ENERGY MODELED BY MCNP5Health Physics, 2007
- MONTE CARLO SIMULATION OF A SCANNING DETECTOR WHOLE BODY COUNTER AND THE EFFECT OF BOMAB PHANTOM SIZE ON THE CALIBRATIONHealth Physics, 2002
- Experiences in Calibration with Three Different Types of PhantomsRadiation Protection Dosimetry, 2000
- Metrological Parameters of the Unified Calibration Whole-Body Phantom with Gamma-Emitting RadionuclidesRadiation Protection Dosimetry, 2000
- The Canadian National Calibration Reference Center for Bioassay and In-vivo MonitoringHealth Physics, 1994
- Total body potassium calibrations for normal and obese subjects in two types of whole body counterPhysics in Medicine & Biology, 1979