Effects of B1 inhomogeneity correction for three‐dimensional variable flip angle T1 measurements in hip dGEMRIC at 3 T and 1.5 T
Open Access
- 29 August 2011
- journal article
- other
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 67 (6) , 1776-1781
Abstract
Delayed gadolinium‐enhanced MRI of cartilage is a technique for studying the development of osteoarthritis using quantitative T1 measurements. Three‐dimensional variable flip angle is a promising method for performing such measurements rapidly, by using two successive spoiled gradient echo sequences with different excitation pulse flip angles. However, the three‐dimensional variable flip angle method is very sensitive to inhomogeneities in the transmitted B1 field in vivo. In this study, a method for correcting for such inhomogeneities, using an additional B1 mapping spin‐echo sequence, was evaluated. Phantom studies concluded that three‐dimensional variable flip angle with B1 correction calculates accurate T1 values also in areas with high B1 deviation. Retrospective analysis of in vivo hip delayed gadolinium‐enhanced MRI of cartilage data from 40 subjects showed the difference between three‐dimensional variable flip angle with and without B1 correction to be generally two to three times higher at 3 T than at 1.5 T. In conclusion, the B1 variations should always be taken into account, both at 1.5 T and at 3 T. Magn Reson Med, 2011.Keywords
This publication has 16 references indexed in Scilit:
- Repeatability of T1‐quantification in dGEMRIC for three different acquisition techniques: Two‐dimensional inversion recovery, three‐dimensional look locker, and three‐dimensional variable flip angleJournal of Magnetic Resonance Imaging, 2010
- High resolution fast T1 mapping technique for dGEMRICJournal of Magnetic Resonance Imaging, 2009
- Comparison of delayed gadolinium enhanced MRI of cartilage (dGEMRIC) using inversion recovery and fast T1 mapping sequencesMagnetic Resonance in Medicine, 2008
- Rapid 3D‐T1 mapping of cartilage with variable flip angle and parallel imaging at 3.0TJournal of Magnetic Resonance Imaging, 2007
- T1 Measurements incorporating flip angle calibration and correction in vivoJournal of Magnetic Resonance, 2006
- dGEMRIC (delayed gadolinium‐enhanced MRI of cartilage) indicates adaptive capacity of human knee cartilageMagnetic Resonance in Medicine, 2004
- Double inversion black‐blood fast spin‐echo imaging of the human heart: A comparison between 1.5T and 3.0TJournal of Magnetic Resonance Imaging, 2003
- Imaging of the active B1 field in vivoMagnetic Resonance in Medicine, 1996
- A method for correctly setting the rf flip angleMagnetic Resonance in Medicine, 1989
- Optimizing the precision in T1 relaxation estimation using limited flip anglesMagnetic Resonance in Medicine, 1987