Effects of B1 inhomogeneity correction for three‐dimensional variable flip angle T1 measurements in hip dGEMRIC at 3 T and 1.5 T

Authors:

Abstract
Delayed gadolinium‐enhanced MRI of cartilage is a technique for studying the development of osteoarthritis using quantitative T1 measurements. Three‐dimensional variable flip angle is a promising method for performing such measurements rapidly, by using two successive spoiled gradient echo sequences with different excitation pulse flip angles. However, the three‐dimensional variable flip angle method is very sensitive to inhomogeneities in the transmitted B1 field in vivo. In this study, a method for correcting for such inhomogeneities, using an additional B1 mapping spin‐echo sequence, was evaluated. Phantom studies concluded that three‐dimensional variable flip angle with B1 correction calculates accurate T1 values also in areas with high B1 deviation. Retrospective analysis of in vivo hip delayed gadolinium‐enhanced MRI of cartilage data from 40 subjects showed the difference between three‐dimensional variable flip angle with and without B1 correction to be generally two to three times higher at 3 T than at 1.5 T. In conclusion, the B1 variations should always be taken into account, both at 1.5 T and at 3 T. Magn Reson Med, 2011.