Haemodynamic Monitoring of Cardiac Status Using Heart Sounds from an Implanted Cardiac Device
Open Access
- 4 July 2017
- journal article
- research article
- Published by Oxford University Press (OUP) in ESC Heart Failure
- Vol. 4 (4) , 605-613
Abstract
The aim of this study was to evaluate the haemodynamic correlates of heart sound (HS) parameters such as third HS (S3), first HS (S1), and HS-based systolic time intervals (HSTIs) from an implantable cardiac device. Two unique animal models (10 swine with myocardial ischaemia and 11 canines with pulmonary oedema) were used to evaluate haemodynamic correlates of S1, S3, and HSTIs, namely, HS-based pre-ejection period (HSPEP), HS-based ejection time (HSET), and the ratio HSPEP/HSET during acute haemodynamic perturbations. The HS was measured using implanted cardiac resynchronization therapy defibrillator devices simultaneously with haemodynamic references such as left atrial (LA) pressure and left ventricular (LV) pressure. In the ischaemia model, S1 amplitude (r = 0.76 ± 0.038; P = 0.002), HSPEP (r = −0.56 ± 0.07; P = 0.002), and HSPEP/HSET (r = −0.42 ± 0.1; P = 0.002) were significantly correlated with LV dP/dtmax. In contrast, HSET was poorly correlated with LV dP/dtmax (r = 0.14 ± 0.14; P = 0.23). In the oedema model, a physiological delayed response was observed in S3 amplitude after acute haemodynamic perturbations. After adjusting for the delay, S3 amplitude significantly correlated with LA pressure in individual animals (r = 0.71 ± 0.07; max: 0.92; min: 0.17) as well as in aggregate (r = 0.62; P < 0.001). The S3 amplitude was able to detect elevated LA pressure, defined as >25 mmHg, with a sensitivity = 58% and specificity = 90%. The HS parameters such as S1, S3, and HSTIs measured using implantable devices significantly correlated with haemodynamic changes in acute animal models, suggesting potential utility for remote heart failure patient monitoring.Keywords
Funding Information
- Boston Scientific
- Boston Scientific Corporation
This publication has 36 references indexed in Scilit:
- Feasibility of Heart Sounds Measurements from an Accelerometer within an ICD Pulse GeneratorPacing and Clinical Electrophysiology, 2012
- Endocardial acceleration (sonR) vs. ultrasound-derived time intervals in recipients of cardiac resynchronization therapy systemsEP Europace, 2011
- Hemodynamic Correlates of the Third Heart Sound and Systolic Time IntervalsCongestive Heart Failure, 2006
- Hemodynamic Correlates of the Third Heart Sound and Systolic Time IntervalsCongestive Heart Failure, 2006
- Third heart sound: genesis and clinical importanceInternational Journal of Cardiology, 2004
- Diastolic properties of the left ventricle in normal adults and in patients with third heart sounds.Circulation, 1984
- Origin of the third heart sound. I. Studies in dogs.Circulation, 1983
- A New Inotropic Drug: Its Promise and a CautionNew England Journal of Medicine, 1978
- A critical review of the systolic time intervals.Circulation, 1977
- Systolic Time Intervals as Measures of the Contractile State of the Left Ventricular Myocardium in ManCirculation, 1972