Development and validation of a 1.5 T MR‐Linac full accelerator head and cryostat model for Monte Carlo dose simulations
Authors:
Open Access
- 18 September 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 46 (11) , 5304-5313
Abstract
Purpose To develop, implement and validate a full 1.5 T/7 MV MR‐Linac accelerator head and cryostat model in EGSnrc for high precision dose calculations accounting for magnetic field effects that are independent from the vendor treatment planning system. Methods Primary electron beam parameters for the implemented model were adapted to be in accordance with measured dose profiles of the Elekta Unity (Elekta AB, Stockholm, Sweden). Parameters to be investigated were the mean electron energy as well as the gaussian radial intensity and energy distributions. Energy tuning was done comparing depth dose profiles simulated with monoenergetic beams of varying energies to measurements. The optimum radial intensity distribution was found by varying the radial FWHM and comparing simulated and measured lateral profiles. The influence of the energy distribution was investigated comparing simulated lateral and depth dose profiles with varying energy spreads to measured data. Comparison of simulations and measurements was performed by calculating average and maximum local dose deviations. The model was validated recalculating a clinical IMRT plan for the MR‐Linac and comparing the resulting dose distribution with simulations from the commercial treatment planning system Monaco using the gamma criterion. Results Comparison of simulated and measured data showed that the optimum initial electron beam for MR‐Linac simulations was monoenergetic with an electron energy of (7.4 ± 0.2) MeV. The optimum gaussian radial intensity distribution has a FWHM of (2.2 ± 0.3) mm. Average relative deviations were smaller than 1% for all simulated profiles with optimum electron parameters, whereas the largest maximum deviation of 2.07% was found for the 22 × 22 cm2 cross plane profile. Profiles were insensitive to energy spread variations. The IMRT plan recalculated with the final MR‐Linac model with optimized initial electron beam parameters showed a gamma pass rate of 99.83% using a gamma criterion of 3%/3mm. Conclusions The EGSnrc MR‐Linac model developed in this study showed good accordance with measurements and was successfully used to recalculate a first full clinical IMRT treatment plan. Thus, it shows the general possibility for future secondary dose calculations of full IMRT plans with EGSnrc, which needs further detailed investigations before clinical use.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (INST 39/963‐1)
This publication has 27 references indexed in Scilit:
- Adaptive radiotherapy: The Elekta Unity MR-linac conceptClinical and Translational Radiation Oncology, 2019
- Ionization chamber correction factors for MR-linacsPhysics in Medicine & Biology, 2018
- Magnetic Resonance Imaging-Guided Adaptive Radiation Therapy: A “Game Changer” for Prostate Treatment?International Journal of Radiation Oncology - Biology - Physics (IJROBP), 2018
- Potential Applications of Image-Guided Radiotherapy for Radiation Dose Escalation in Patients with Early Stage High-Risk Prostate CancerFrontiers in Oncology, 2015
- MR guidance in radiotherapyPhysics in Medicine & Biology, 2014
- Introduction: Systems for Magnetic Resonance Image Guided Radiation TherapySeminars in Radiation Oncology, 2014
- The Magnetic Resonance Imaging–Linac SystemSeminars in Radiation Oncology, 2014
- MRI/linac integrationRadiotherapy and Oncology, 2008
- Dose optimization for the MRI-accelerator: IMRT in the presence of a magnetic fieldPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005