Abstract
Bone preservation and physiological distribution of forces on the proximal femur are key elements in introducing a successful uncemented total hip replacement. In order to achieve this, in the mid 1990s, we developed an ultra short proximal loading custom-made component with a lateral flare, a high femoral neck osteotomy and without a diaphyseal stem. We report the outcome of 129 custom-made hydroxyapatite-coated uncemented short femoral components inserted into 109 patients between June 1995 and May 2004. The mean age of the patients was 51 years (21 to 71) and the mean follow-up was eight years (4.9 to 14.1). Bone behaviour around the implant was studied on the post-operative radiographs. The mean Harris hip score improved from 44 (8 to 66) pre-operatively to 95 (76 to 100) at final follow-up. The Western Ontario MacMaster University Osteoarthritis index was 93 of 100 at final review. None of the patients reported thigh pain. A total of five hips were revised, three for polyethylene liner exchange and two for complete revision of the acetabular component. No femoral components were revised. The radiological changes in the proximal femur were generally good, as evidenced by spot welds both on the medial and lateral aspects of the femur. No component migrated. The presence of a lateral flare and use of a high osteotomy of the femoral neck provided good clinical and radiological results. The absence of a diaphyseal portion of the stem did not impair stability.Keywords
This publication has 28 references indexed in Scilit:
- An in vitro study of the strain distribution in human femora with anatomical and customised femoral stemsThe Journal of Bone and Joint Surgery. British volume, 2009
- Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stemsActa Orthopaedica, 2009
- Perspectives on metaphyseal conservative stemsJournal of Orthopaedics and Traumatology, 2008
- New Femoral DesignsClinical Orthopaedics and Related Research, 2006
- Seven to 10 Years Followup of an Anatomic Hip ProsthesisClinical Orthopaedics and Related Research, 2004
- Closeness of fit of uncemented stems improves the strain distribution in the femurJournal of Orthopaedic Research, 1995
- Fit of the uncemented femoral component and the use of cement influence the strain transfer to the femoral cortexJournal of Orthopaedic Research, 1994
- The contribution of the nonporous distal stem to the stability of proximally porous-coated canine femoral componentsThe Journal of Arthroplasty, 1993
- Determinants of Stress ShieldingClinical Orthopaedics and Related Research, 1992
- "Modes of Failure" of Cemented Stem-type Femoral ComponentsClinical Orthopaedics and Related Research, 1979