Comparative safety of stent retriever designs for intracranial atherosclerotic disease related large vessel occlusion using agarose models
Open Access
- 20 November 2025
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 15 (1) , 40952
Abstract
Establishing an optimal first-line approach for suspected intracranial atherosclerotic disease (ICAD)-related acute large vessel occlusion (LVO) remains challenging. We developed an ICAD model using an agarose phantom to investigate the specific features of the stent retriever (SR) design for ICAD-LVO for safety use. An ICAD model that reproduced the mechanical features of plaque was developed to quantify the pull-out resistance and intraluminal injury caused by SRs. The impacts of vessel factors (ICAD existence and plaque stiffness) and SR design (non-segmented, segmented, and manually-controllable-diameter) were evaluated using this model. In the 6% agarose (as soft plaque) ICAD model, SR caused significantly higher pull-out resistance and severe intraluminal injury compared with the non-ICAD model or the 10% agarose (as hard plaque) ICAD model with almost all SRs. In the 6% ICAD model, non-segmented, rather than segmented, SRs seemed to reduce pull-out resistance and intraluminal injury. Longer SRs strengthened the benefit of non-segmented design. SRs with manually controllable diameter, when fully relaxed before retrieval, seemed to provide the safest option among all SR designs, while those with usual handling such as slight relaxation before retrieval seemed to be the most harmful due to specific deformation. Using SRs for ICAD-LVO may increase pull-out resistance and intraluminal injury, particularly in ICAD involving soft plaques. SR use safety for ICAD seems to depend largely on SR design and handling procedures of manually controllable diameter SRs.Keywords
Funding Information
- the Japan Agency for Medical Research and Development (JP19ek0210088, JP20ek0210129, JP20ek0210147, JP21ek0210147, JP22ek0210147)
- The Japanese Ministry of Health, Labour and Welfare (H28-Shinkin-Ippan-011, 19AC1003, 21FA1010, 22FA1015, 23FA1014, 24FA1015, 24FA1016)
- Japan Society for the Promotion of Science (25293314, 18H02914, 22H03191, 23K24450)
- Intramural Research Fund (20-4-10)
This publication has 33 references indexed in Scilit:
- Clinico-radiological features of intracranial atherosclerosis-related large vessel occlusion prior to endovascular treatmentScientific Reports, 2024
- Balloon angioplasty as first-choice recanalization strategy for intracranial atherosclerosis-related emergent large vessel occlusion with small clot burdenNeuroradiology, 2024
- Tigertriever in the treatment of acute ischemic stroke with underlying intracranial atherosclerotic diseaseJournal of NeuroInterventional Surgery, 2023
- Functional Tough Hydrogels: Design, Processing, and Biomedical ApplicationsAccounts of Materials Research, 2022
- Hypoperfusion intensity ratio and CBV index as predictive parameters to identify underlying intracranial atherosclerotic stenosis in endovascular thrombectomyJournal Of Neuroradiology, 2022
- Experimental evaluation of stent retrievers’ mechanical properties and effectivenessJournal of NeuroInterventional Surgery, 2016
- Histological examination of vascular damage caused by stent retriever thrombectomy devicesJournal of NeuroInterventional Surgery, 2015
- Endovascular Therapy of Cerebral Arterial Occlusions: Intracranial Atherosclerosis versus EmbolismJournal of Stroke and Cerebrovascular Diseases, 2015
- Quantitative Measurement of Dissection Resistance in Intimal and Medial Layers of Human Coronary ArteriesExperimental Mechanics, 2013
- Experimental measurement of the mechanical properties of carotid atherothrombotic plaque fibrous capJournal of Biomechanics, 2009