Identifying and Correcting Scan Noise and Drift in the Scanning Transmission Electron Microscope
Authors:
- 14 May 2013
- journal article
- materials applications
- Published by Oxford University Press (OUP) in Microscopy and Microanalysis
- Vol. 19 (4) , 1050-1060
Abstract
The aberration-corrected scanning transmission electron microscope has great sensitivity to environmental or instrumental disturbances such as acoustic, mechanical, or electromagnetic interference. This interference can introduce distortions to the images recorded and degrade both signal noise and resolution performance. In addition, sample or stage drift can cause the images to appear warped and leads to unreliable lattice parameters being exhibited. Here a detailed study of the sources, natures, and effects of imaging distortions is presented, and from this analysis a piece of image reconstruction code has been developed that can restore the majority of the effects of these detrimental image distortions for atomic-resolution data. Example data are presented, and the performance of the restored images is compared quantitatively against the as-recorded data. An improvement in apparent resolution of 16% and an improvement in signal-to-noise ratio of 30% were achieved, as well as correction of the drift up to the precision to which it can be measured.Keywords
This publication has 16 references indexed in Scilit:
- Correcting scanning instabilities from images of periodic structuresUltramicroscopy, 2012
- The Structure of Grain Boundaries in Strontium Titanate: Theory, Simulation, and Electron MicroscopyAnnual Review of Materials Research, 2010
- Aberration-corrected scanning transmission electron microscopy: from atomic imaging and analysis to solving energy problemsPhilosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences, 2009
- Development of a stage-scanning system for high-resolution confocal STEMJournal of Electron Microscopy, 2008
- Room design for high-performance electron microscopyUltramicroscopy, 2006
- IMAGE-WARP: A real-space restoration method for high-resolution STEM images using quantitative HRTEM analysisUltramicroscopy, 2005
- Fluctuation microscopy in the STEMUltramicroscopy, 2002
- Upper limits for the residual aberrations of a high-resolution aberration-corrected STEMUltramicroscopy, 2000
- Towards sub-Å electron beamsUltramicroscopy, 1999
- The Sensitivity Performance of the Human Eye on an Absolute Scale*Journal of the Optical Society of America, 1948