Design, Fabrication, and Operation of a Nitrogen Monoxide Measurement Device Based on LTCC

Authors:

Abstract
We present a so-called micro total analytical system (μTAS) for the measurement of low nitrogen monoxide (NO) concentrations in gas flows. The measurement principle is based on the chemiluminescent reaction of NO with ozone (O3). The presented system scales down this measurement principle into the realm of microfluidics and microsensors. Low temperature cofired ceramics (LTCC) have proven to be the ideal technology for the realization of a μTAS, as they offer high chemical and thermal stability as well as high degree of freedom of design. The system consists of four components: an ozone generator, an ozone concentration measurement via UV transmission, a chemiluminescent detection (CLD) chamber, and an exhaust gas treatment. The ozone generator is a microplasma device in which ozone is produced by a dielectric barrier discharge (DBD) mechanism. From the generator, the ozone is fed into a chamber with an optical port, where the concentration is determined by UV transmission measurements. Then it is brought into contact with the NO-containing analyte gas inside the reaction chamber, where the chemiluminescent reaction occurs. The reaction chamber is outfitted with an optical port as well, so that the emitted radiation can be detected and analyzed with a photodiode. The gas mixture leaving the reaction chamber is fed through a platinum catalyst, in order to decompose leftover ozone and prevent buildup of hazardous concentrations in the surroundings of the device. This paper presents the design of each component as well as the integrated system and gives detailed measurements of its operation.