Cyclic polyamide oligomers extracted from nylon 66 membrane filter disks as a source of contamination in liquid chromatography/mass spectrometry
Authors:
- 1 May 2006
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 17 (5) , 652-656
Abstract
Background interferences are perhaps an unavoidable part of analytical detection schemes, particularly when analyzing trace level samples or when using detectors with high levels of sensitivity. In liquid chromatography, mobile phase solvents are routinely filtered using membrane filter disks to trap particulates in hopes of minimizing contamination, providing improvements in data output and instrumental operation. In this study, we report that one such filter disk leads to a significant level of contamination in LC and LC/MS experiments. Extractable compounds from nylon membrane filters generate significant background signals in UV absorption chromatograms at 214 nm, and are also detected by electrospray ionization mass spectrometry, with nominal m/z values of 453 and 679. It is shown that rinsing the nylon membranes before their use can reduce, but will not eliminate, the extractable contaminants from the mobile phase. Through MS and tandem MS analysis, we have identified these contaminants as cyclic oligomers of polyamide 66. Based on these results, it is recommended that nylon membrane filters be avoided when conducting trace level analysis, particularly when conducting LC/MS experiments.Keywords
This publication has 24 references indexed in Scilit:
- Ghost peaks in reversed-phase gradient HPLC: a review and updateJournal of Chromatography A, 2004
- Ultrapure water for liquid chromatography–mass spectrometry studiesJournal of Chromatography A, 2004
- Volatile organic components migrating from plastic pipes (HDPE, PEX and PVC) into drinking waterWater Research, 2003
- On the mature of the chemical noise in MALDI mass spectraJournal of the American Society for Mass Spectrometry, 2002
- A combined linear ion trap time‐of‐flight system with improved performance and MS n capabilitiesRapid Communications in Mass Spectrometry, 2001
- The effect of using silicon based diffusion pump fluid on spectral quality in an electrospray ionization ion trap/time-of-flight mass spectrometerReview of Scientific Instruments, 1997
- Interferences in solid-phase extraction using C-18 bonded porous silica cartridgesAnalytical Chemistry, 1988
- Anomalous liquid chromatographic results originating from the presence of soap residues and the use of nylon filtersJournal of Chromatography A, 1987
- Degradation of High Purity Water on StorageJournal of Liquid Chromatography, 1983
- Detection of trace organic impurities in binary solvent systems: A solvent purity testJournal of Chromatography A, 1980