Lung assist devices influence cardio-energetic parameters: Numerical simulation study
- 1 August 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 124 (1094687X) , 4515-4519
Abstract
We aim at an analysis of the effects mechanical ventilators (MVs) and thoracic artificial lungs (TALs) will have on the cardiovascular system, especially on important quantities, such as left and right ventricular external work (EW), pressure-volume area (PVA) and cardiac mechanical efficiency (CME). Our analyses are based on simulation studies which were carried out by using our CARDIOSIM © software simulator. At first, we carried out simulation studies of patients undergoing mechanical ventilation (MV) without a thoracic artificial lung (TAL). Subsequently, we conducted simulation studies of patients who had been provided with a TAL, but did not undergo MV. We aimed at describing the patient's physiological characteristics and their variations with time, such as EW, PVA, CME, cardiac output (CO) and mean pulmonary arterial/venous pressure (PAP/PVP). We were starting with a simulation run under well-defined initial conditions which was followed by simulation runs for a wide range of mean intrathoracic pressure settings. Our simulations of MV without TAL showed that for mean intrathoracic pressure settings from negative (-4 mmHg) to positive (+5 mmHg) values, the left and right ventricular EW and PVA, right ventricular CME and CO decreased, whereas left ventricular CME and the PAP increased. The simulation studies of patients with a TAL, comprised all the usual TAL arrangements, viz. configurations “in series” and in parallel with the natural lung and, moreover, hybrid configurations. The main objective of the simulation studies was, as before, the assessment of the hemodynamic response to the application of a TAL. We could for instance show that, in case of an “in series” configuration, a reduction (an increase) in left (right) ventricular EW and PVA values occurred, whereas the best performance in terms of CO can be achieved in the case of an in parallel configuration.Keywords
This publication has 9 references indexed in Scilit:
- Mechanical ventilation and thoracic artificial lung assistance during mechanical circulatory support with PUCA pump: In silico studyComputer Methods and Programs in Biomedicine, 2013
- Review of Zero-D and 1-D Models of Blood Flow in the Cardiovascular SystemBioMedical Engineering OnLine, 2011
- Computer Simulation of Coronary Flow Waveforms during Caval OcclusionMethods of Information in Medicine, 2009
- Modelling in the study of interaction of Hemopump device and artificial ventilationComputers in Biology and Medicine, 2006
- In Vivo and Simulation Study of Artificial Ventilation Effects on Energetic Variables in Cardiosurgical PatientsMethods of Information in Medicine, 2005
- Predictors of Severe Right Ventricular Failure After Implantable Left Ventricular Assist Device Insertion: Analysis of 245 PatientsCirculation, 2002
- Hemodynamic Effects of Attachment Modes and Device Design of a Thoracic Artificial LungASAIO Journal, 2000
- A modular numerical model of the cardiovascular system for studying and training in the field of cardiovascular physiopathologyJournal of Biomedical Engineering, 1992
- Ventricular systolic interdependence: volume elastance model in isolated canine heartsAmerican Journal of Physiology-Heart and Circulatory Physiology, 1987