A 3-D PW ultrasonic Doppler flowmeter: theory and experimental characterization
Authors:
- L. CapineriL. Capineri
- A. FortA. Fort
- L. MasottiLML. Masotti
- S. RocchiSRS. Rocchi
- M. ScabiaMSM. Scabia
- 1 January 1999
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 46 (1) , 108-113
Abstract
A complete 3-D ultrasonic pulsed Doppler system has been developed to measure quantitatively the velocity vector field of a fluid flow independently of the probe position. The probe consists of four 2.5 MHz piezocomposite ultrasonic transducers (one central transmitter and three receivers separated by 120/spl deg/) to measure the velocity projections along three different directions. The Doppler shift of the three channels is calculated by analog phase and quadrature demodulation, then digitally processed to extract the mean velocity from the complex spectrum. The accuracy of the 3-D Doppler technique has been tested on a moving string phantom providing an error of about 4% for both amplitude and direction with an acquisition window of 100 ms.Keywords
This publication has 11 references indexed in Scilit:
- Triple-beam lens transducers for three-dimensional ultrasonic fluid flow estimationIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1995
- Vector doppler: Accurate measurement of blood velocity in two dimensionsUltrasound in Medicine & Biology, 1992
- A flexible blood flow phantom capable of independently producing constant and pulsatile flow with a predictable spatial flow profile for ultrasound flow measurement validationsIEEE Transactions on Biomedical Engineering, 1992
- 3-D velocity field measurement using multiple ultrasonic plane detections and high-order correlation analysisIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1991
- A nonstationary signal simulation model for continuous wave and pulsed Doppler ultrasoundIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1989
- Three-dimensional Doppler velocimetry of flow jetsIEEE Transactions on Biomedical Engineering, 1988
- Angle Independent Ultrasonic Detection of Blood FlowIEEE Transactions on Biomedical Engineering, 1987
- Ultrasound Doppler Probing of Flows Transverse with Respect to Beam AxisIEEE Transactions on Biomedical Engineering, 1987
- Time Dilation and Inversion Properties and the Output Spectrum of Pulsed Doppler FlowmetersIEEE Transactions on Sonics and Ultrasonics, 1983
- Evaluating doppler devices using a moving string test targetJournal of Clinical Ultrasound, 1982