Linear Accuracy of Cone Beam CT Derived 3D Images
- 1 January 2009
- journal article
- research article
- Published by The Angle Orthodontist (EH Angle Education & Research Foundation) in The Angle Orthodontist
- Vol. 79 (1) , 150-157
Abstract
Objective: To compare the in vitro reliability and accuracy of linear measurements between cephalometric landmarks on cone beam computed tomography (CBCT) 3D volumetric images with varying basis projection images to direct measurements on human skulls. Materials and Methods: Sixteen linear dimensions between 24 anatomic sites marked on 19 human skulls were directly measured. The skulls were imaged with CBCT (i-CAT, Imaging Sciences International, Hatfield, Pa) at three settings: (a) 153 projections, (b) 306 projections, and (c) 612 projections. The mean absolute error and modality mean (± SD) of linear measurements between landmarks on volumetric renderings were compared to the anatomic truth using repeated measures general linear model (P ≤ .05). Results: No difference in mean absolute error between the scan settings was found for almost all measurements. The average skull absolute error between marked reference points was less than the distances between unmarked reference sites. CBCT resulted in lower measurements for nine dimensions (mean difference range: 3.1 mm ± 0.12 mm to 0.56 mm ± 0.07 mm) and a greater measurement for one dimension (mean difference 3.3 mm ± 0.12 mm). No differences were detected between CBCT scan sequences. Conclusions: CBCT measurements were consistent between scan sequences and for direct measurements between marked reference points. Reducing the number of projections for 3D reconstruction did not lead to reduced dimensional accuracy and potentially provides reduced patient radiation exposure. Because the fiducial landmarks on the skulls were not radio-opaque, the inaccuracies found in measurement could be due to the methods applied rather than to innate inaccuracies in the CBCT scan reconstructions or 3D software employed.Keywords
This publication has 16 references indexed in Scilit:
- Accuracy of linear measurements from imaging plate and lateral cephalometric images derived from cone-beam computed tomographyAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2007
- Comparison of conventional and cone beam CT synthesized cephalogramsDentomaxillofacial Radiology, 2007
- Development of imaging selection criteria and procedures should precede cephalometric assessment with cone-beam computed tomographyAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2006
- Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CATDentomaxillofacial Radiology, 2006
- Accuracy of linear temporomandibular joint measurements with cone beam computed tomography and digital cephalometric radiographyAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2005
- Dose reduction in maxillofacial imaging using low dose Cone Beam CTEuropean Journal of Radiology, 2005
- Proposed reference point for 3-dimensional cephalometric analysis with cone-beam computerized tomographyAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2005
- Radiation exposure during midfacial imaging using 4- and 16-slice computed tomography, cone beam computed tomography systems and conventional radiographyDentomaxillofacial Radiology, 2004
- Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unitDentomaxillofacial Radiology, 2003
- A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machineOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2003