Determination of manganese‐ and manganese‐containing fungicides with lucigenin–Tween‐20‐enhanced chemiluminescence detection
Authors:
- 30 January 2015
- journal article
- research article
- Published by Wiley in Luminescence
- Vol. 30 (7) , 950-961
Abstract
A flow‐injection (FI) method is reported for the determination of Mn(II), maneb and mancozeb fungicides based on the catalytic effect of Mn(II) on the oxidation of lucigenin and dissolved oxygen in a basic solution. The Tween‐20 surfactant has been reported for first time to enhance lucigenin chemiluminescence (CL) intensity in the presence of Mn(II) (53%) and maneb and mancozeb (89%). The calibration graphs were linear in the concentration range of 0.001–1.5 mg L–1 (R2 = 0.9982 (n = 11) with a limit of detection (S/N = 3) of 0.1 µg L–1 for Mn(II) and 0.01–3.0 mg L–1 [R2 = 0.9989 and R2 = 0.9992 (n = 6)] with a limit of detection (S/N =3) of 1.0 µg L–1 for maneb and mancozeb respectively. Injection throughputs of 90 and 120 h–1 for Mn(II) and maneb and mancozeb respectively, and relative standard deviations of 1.0–3.4% were obtained in the concentration range studied. The experimental variables, e.g., reagents concentrations, flow rates, sample volume, and photomultiplier tube voltage, were optimized and potential interferences were investigated. The analysis of Mn(II) in river water reference materials (SLRS‐4 and SLRS‐5) showed good agreement with the certified values incorporating an on‐line 8‐hydroxyquinoline chelating column in the manifold for removing interfering metal ions. Recoveries for maneb and mancozeb were in the range of 92 ± 5 to 104 ± 3% and 91 ± 2 to 100 ± 4% (n = 3) respectively. The effect of 30 other pesticides (fungicides, herbicides and insecticides) was also examined in the lucigenin–Tween‐20 CL system. Copyright © 2015 John Wiley & Sons, Ltd.Keywords
Funding Information
- Indigenous PhD Scholarship (PIN 117-6914-PS7-156)
- Department of Chemistry
This publication has 55 references indexed in Scilit:
- Colorimetric determination of manganese(II) in natural and waste waters after preconcentration on a polymeric adsorbentJournal of Analytical Chemistry, 2011
- Dissolved manganese in the Atlantic sector of the Southern OceanDeep Sea Research Part II: Topical Studies in Oceanography, 2011
- Investigation of iron(III) reduction and trace metal interferences in the determination of dissolved iron in seawater using flow injection with luminol chemiluminescence detectionAnalytica Chimica Acta, 2009
- Determination of trace amounts of manganese in natural waters by flow injection stopped-flow catalytic kinetic spectrophotometryAnalytica Chimica Acta, 2004
- Chemiluminescence of dithiocarbamate fungicides based on the luminol reactionAnalytica Chimica Acta, 2003
- Preconcentration of manganese(II) from natural and sea water on a palmitoyl quinolin-8-ol functionalized XAD copolymer resin and spectrophotometric determination with the formaldoxime reagentAnalytica Chimica Acta, 2003
- Selective and sensitive determination of trace manganese in sea water by flow through technique using luminol–hydrogen peroxide chemiluminescence detectionAnalytica Chimica Acta, 1998
- Determination of trace elements in seawater by fluorinated metal alkoxide glass-immobilized 8-hydroxyquinoline concentration and high-resolution inductively coupled plasma mass spectrometry detectionAnalytica Chimica Acta, 1998
- Determination of manganese in natural water and effluent streams using a solid-phase lead(IV) dioxide reactor in a flow-injection systemAnalytica Chimica Acta, 1997
- Determination of dissolved manganese(II) in coastal and estuarine waters by differential pulse cathodic stripping voltammetryAnalytica Chimica Acta, 1997