Valiant Thoracic Stent-Graft Deployed With the New Captivia Delivery System:Procedural and 30-Day Results of the Valiant Captivia Registry
- 1 April 2012
- journal article
- research article
- Published by International Society of Endovascular Specialists in Journal of Endovascular Therapy
- Vol. 19 (2) , 213-225
Abstract
Purpose To evaluate procedural and 30-day outcomes of thoracic endovascular aortic repair (TEVAR) employing the Valiant Thoracic Stent Graft with the Captivia Delivery System. Methods Enrollment in the study (www.ClinicalTrials.com identifier NCT01181947) included all eligible patients implanted with the Valiant Captivia System retrospectively and prospectively at 15 sites in Europe and Turkey between October 2009 and June 2010. In the 100 treated patients (81 men; mean age 64.6±12.0 years, range 25–87), indications included descending thoracic aortic aneurysm (TAA, 49.0%) and aortic dissection (42.0%). Results Technical success was 100.0%, with no misaligned deployments or aortic perforations. Mean follow-up was 68.9±34.9 days (range 20–147, median 61). The 30-day rate of all-cause mortality was 4.0% (all 4 cases procedure-related, 3 device-related). Retrograde type A dissection occurred in 2 patients. The only conversion to open surgery was successful in a patient experiencing intraoperative aneurysm rupture. Stroke occurred in 4 (4.0%) patients and paraplegia in 1 (1.0%). Among 66 patients with 30-day imaging studies evaluable for endoleak, 4 (6.1%) had type I and 7 (10.6%) had type II endoleak; there were no types III or IV. Within 30 days, no secondary endovascular procedures were required due to endoleak. One patient with type II endoleak died 3 weeks postimplantation before scheduled embolization. Conclusion In this analysis of procedural and 30-day results, the high technical success and clinical outcome rates showed that the Valiant Thoracic Stent Graft with the new Captivia Delivery System has promising capacity to treat a variety of thoracic aortic conditions in a range of anatomies.Keywords
This publication has 13 references indexed in Scilit:
- The VIRTUE Registry of Type B Thoracic Dissections – Study Design and Early ResultsEuropean Journal of Vascular and Endovascular Surgery, 2011
- Incidence of and outcomes after misaligned deployment of the Talent Thoracic Stent Graft SystemJournal of Vascular Surgery, 2010
- Midterm Results From the TRAVIATA Registry: Treatment of Thoracic Aortic Disease With the Valiant Stent GraftJournal of Endovascular Therapy, 2010
- Commentary: The Valiant Thoracic Stent Graft for Endovascular Repair of Thoracic Aortic Aneurysms and Dissections: Looking Good So FarJournal of Endovascular Therapy, 2010
- Endovascular Aortic Repair Versus Open Surgical Repair for Descending Thoracic Aortic DiseaseJACC, 2010
- Endovascular Repair of Descending Thoracic Aortic Aneurysm: Review of LiteratureSeminars in Thoracic and Cardiovascular Surgery, 2010
- Failure modes of thoracic endografts: Prevention and managementJournal of Vascular Surgery, 2009
- Expert Consensus Document on the Treatment of Descending Thoracic Aortic Disease Using Endovascular Stent-GraftsThe Annals of Thoracic Surgery, 2008
- Early Results of Endovascular Treatment of the Thoracic Aorta Using the Valiant EndograftCardioVascular and Interventional Radiology, 2007
- Perioperative differences between endovascular repair of thoracic and abdominal aortic diseasesJournal of Vascular Surgery, 2007