Sample Preconcentration Using Ion-Exchange Polymer Film for Laser Ablation Sampling Inductively Coupled Plasma Atomic Emission Spectrometry
Authors:
- 1 December 1997
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 69 (23) , 4872-4877
Abstract
An efficient sample pretreatment/introduction technique for the inductively coupled plasma atomic emission spectrometry (ICP-AES) using ion exchange for analyte preconcentration and matrix separation and laser ablation sampling for sample introduction has been developed. Ammonium pyrrolidine dithiocarbamate (APDC)−polystyrene films are coated on glass plates for analyte preconcentration. Repetitive laser ablation sampling of the polymer film removes the ion-exchanged metal ions from the polymer film as fine particles for sample introduction into the ICP. After immersing the sample probe in a sample solution for 5 min, the ICP emission intensity for laser ablation of the polymer film is a few times larger than that after solution nebulization. The sample probe removes only a small fraction of the sample solution and, therefore, in principle, does not disturb the original solution significantly. Single-pulse laser ablation of the polymer film shows that the ion-exchanged metal ion concentration in the film reduces exponentially with the depth of the polymer film. Ion exchange to the polymer film is probably limited by the rate of metal ion diffusion into the film. Calibration curves for Cu, Hg, Pb, and Zn show linear dynamic range of ∼1−2 orders of magnitude. The linear dynamic range for Cu increases to >3 orders of magnitude when using Pb as an internal standard. RSD of the ICP emission intensity is ∼8%.Keywords
This publication has 14 references indexed in Scilit:
- Determination of elements in polymers by laser ablation inductively coupled plasma atomic emission spectrometry: effect of the laser beam wavelength, energy and masking on the ablation threshold and efficiencySpectrochimica Acta Part B: Atomic Spectroscopy, 1996
- Determination of trace amounts of copper, nickel and zinc in palladium compounds by solvent extraction flame atomic absorption spectrometryThe Analyst, 1991
- Chelation ion chromatography as a method for trace elemental analysis in complex environmental and biological samplesAnalytical Chemistry, 1990
- Determination of lead in sheep's blood by graphite furnace atomic absorption spectrometry using a modified chelation and extraction procedureThe Analyst, 1987
- Chelating Sorbents in Analytical ChemistryCritical Reviews in Analytical Chemistry, 1986
- Simultaneous determination of trace metals in sea water using dithiocarbamate pre-concentration and inductively coupled plasma emission spectrometryThe Analyst, 1981
- Concentration and spectrochemical determination of trace metals in urine with a poly(dithiocarbamate) resin and inductively coupled plasma-atomic emission spectrometryAnalytical Chemistry, 1979
- Concentration of heavy metals by complexation on dithiocarbamate resinsAnalytical Chemistry, 1974
- APDC-MIBK extraction system for the determination of trace elements in saline waters by atomic-absorption spectrophotometryTalanta, 1967
- Über die Verwendung von substituierten Dithiocarbamaten in der Mikroanalyse. IIIMicrochimica Acta, 1955