Photon-Counting CT Enables Higher Image Quality at a Lower Radiation Dose Compared with a 256-Slice Energy-Integrating Detector CT in Pediatric Patients with Congenital Heart Disease
Open Access
- 1 March 2026
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 16 (5) , 735
Abstract
Background/Objectives: There is an increasing need for cross-sectional imaging in pediatric patients with congenital heart disease. This patient group is highly sensitive to ionizing radiation. The purpose of this article was to evaluate differences in radiation dose and image quality between a first-generation photon-counting (PC)-CT system and a 256-slice single-source energy-integrating detector (EID)-CT in these patients. Methods: We retrospectively assessed effective dose, CT dose index-volume (CTDIvol), dose length product (DLP), and image quality of all prospectively electrocardiography-gated CTs of the thorax in all patients < 18 years of age examined between February 2021 and August 2024 (n = 43). Two independent observers subjectively scored image quality, vascular contrast, and noise on a 5-point Likert scale. In addition, we assessed the signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) quantitatively. Results: All dose parameters were significantly lower in the PC-CT (n = 27) versus the EID-CT (n = 16) group (mean effective dose: 0.8 ± 0.64 versus 2.2 ± 0.88 mSv, p < 0.001; mean CTDIvol: 1.22 ± 0.96 versus 4.8 ± 1.08 mGy, p < 0.001; mean DLP: 30.7 ± 31.9 versus 73.7 ± 50.7 mGy*cm, p < 0.001). Overall subjective image quality and contrast were rated higher in the PC-CT group (p = 0.046 and < 0.001, respectively). Quantitative CNR was significantly higher in the PC-CT group (mean 39.1 ± 12.9 versus 26.2 ± 10.8, p = 0.002). Conclusions: PC-CT enables high-quality examinations for the evaluation of congenital heart disease with a highly significant dose reduction compared with a 256-slice single-source EID-CT.Keywords
This publication has 31 references indexed in Scilit:
- Reduction of radiation exposure and preserved image quality using photon-counting detector cardiac computed tomography without electrocardiographic gating in children with congenital heart diseaseEuropean Radiology, 2025
- “Feed-and-wrap” technique versus deep sedation for neonatal magnetic resonance imaging: a retrospective comparative studyEuropean Radiology, 2024
- Photon-counting computed tomography for paediatric congenital heart defects yields images of high diagnostic quality with low radiation doses at both 70 kV and 90 kVPediatric Radiology, 2024
- Diagnosis of acute pulmonary embolism: when photon-counting-detector CT replaces energy-integrating-detector CT in daily routineEuropean Radiology, 2024
- Photon-counting CT for diagnosis of acute pulmonary embolism: potential for contrast medium and radiation dose reductionEuropean Radiology, 2023
- Role of CT in Congenital Heart DiseaseCurrent Treatment Options in Cardiovascular Medicine, 2017
- CT and MR imaging in congenital cardiac malformations: Where do we come from and where are we going?Diagnostic and Interventional Imaging, 2016
- Cumulative Radiation Exposure in Pediatric Patients with Congenital Heart DiseasePediatric Cardiology, 2014
- Low-dose prospectively gated 256-slice coronary computed tomographic angiographyThe International Journal of Cardiovascular Imaging, 2009
- The Epidemiology of Cardiovascular Defects, Part I: A Study Based on Data from Three Large Registries of Congenital MalformationsPediatric Cardiology, 2003