Microfluidic microwave biosensor based on biomimetic materials for the quantitative detection of glucose
Open Access
- 24 September 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-11
Abstract
This paper presents a microwave microfluidic biosensor for monitoring blood glucose levels. The glucose sensor is a triple ring microstrip patch antenna integrated with a biomimetic microfluidic device capable of measuring a fixed volume of glucose solution. The sensor was utilized to detect 50–500 mg/dL glucose solutions. The interaction of the glucose solution with the electromagnetic field on the patch's surface influences both the resonance frequency and the magnitude of reflection coefficient. The results indicate that the microfluidic device can reduce experimental error and enhance the correlation between glucose concentration, resonant frequency, and reflection coefficient. Finally, the microfluidic sensor had a sensitivity of 0.25 MHz/(mg/dL), a detection limit as low as 7.7 mg/dL, and correlation coefficients of resonance frequency and reflection coefficient with a glucose concentration of 0.996 and 0.984, respectively. The experiment on the sensor's stability verifies the sensor's excellent stability and rapid response (~ 150 ms). Consequently, the device can be used to differentiate the concentration of glucose solutions, as well as to detect blood glucose levels at an early stage.Keywords
Funding Information
- The Shandong Provincial Natural Science Foundation of China (ZR2018BF027, ZR2021ME052)
- Shandong Province Key Research and Development Program (2019RKB01023)
- National Natural Science Foundation of China (61904092, 62181240278)
This publication has 42 references indexed in Scilit:
- Rapid, Sensitive and Reusable Detection of Glucose by a Robust Radiofrequency Integrated Passive Device Biosensor ChipScientific Reports, 2015
- Classification of diabetes and measurement of blood glucose concentration noninvasively using near infrared spectroscopyInfrared Physics & Technology, 2014
- In Vitro Evaluation of Fluorescence Glucose Biosensor ResponseSensors, 2014
- Broadband Tissue Mimicking Phantoms and a Patch Resonator for Evaluating Noninvasive Monitoring of Blood Glucose LevelsIEEE Transactions on Antennas and Propagation, 2014
- Amperometric Glucose Sensors for Whole Blood Measurement Based on Dehydrogenase Enzymes2012
- An acoustic glucose sensorBiosensors and Bioelectronics, 2012
- Heterogeneous anthropomorphic phantoms with realistic dielectric properties for microwave breast imaging experimentsMicrowave and Optical Technology Letters, 2011
- Dielectric properties of glucose in bulk aqueous solutions: Influence of electrode polarization and modelingBiosensors and Bioelectronics, 2011
- Small-volume solution current-time behavior study for application in reverse iontophoresis-based non-invasive blood glucose monitoringScience China Chemistry, 2010
- Polymer gel dosimetry close to an125I interstitial brachytherapy seedPhysics in Medicine & Biology, 2005