A randomised controlled trial of a new handheld vacuum extraction device
Authors:
Open Access
- 24 July 2005
- journal article
- research article
- Published by Wiley in BJOG: An International Journal of Obstetrics & Gynaecology
- Vol. 112 (11) , 1510-1515
Abstract
Objective To evaluate the effectiveness of a new handheld vacuum delivery device.Design Randomised controlled trial.Setting Southmead Hospital, Bristol, UK.Population One hundred and ninety‐four women requiring delivery with vacuum extraction.Methods Women were recruited into the study while still in early labour. If during delivery, vacuum extraction was indicated they were then randomly allocated to either a new handheld vacuum device, the Kiwi Omnicup, or to a ‘standard’ vacuum cup selected by the obstetrician. The ‘standard group’ consisted of the silastic (66/98, 67%) or the metal cups (32/98, 33%). Study data were recorded at completion of the delivery and at 24–48 hours. An additional brief symptom questionnaire was completed by the mother at 10 days postnatally. Statistical analyses were done on an ‘intention‐to‐treat’ basis.Main outcome measures The primary outcome was the successful completion of delivery with the allocated instrument. Secondary outcomes were substantial fetal scalp trauma and substantial maternal trauma.Results The Omnicup was associated with a significantly higher failure rate than the standard cup [34%vs 21%, odds ratio (OR) = 1.9 (1.01–3.6)]. This difference persisted after adjusting for confounding factors [adjusted OR = 2.25 (1.01–5.0)] and resulted in significantly more forceps deliveries in the Kiwi Omnicup group [22%vs 10%, unadjusted OR = 2.5 (1.1–5.5)]. The caesarean section rates were similar in both groups [9.4%vs 8.2%, OR = 1.2 (0.4–3.2)]. Rates of substantial scalp trauma and maternal trauma were similar in both groups. However, there were more cases of jaundice in the standard cup group [2.3%vs 12.0%, unadjusted OR = 0.18 (0.04–0.82) and adjusted OR = 0.10 (0.01–0.88)].Conclusions The Omnicup is not as successful at achieving a vaginal delivery as the combination of the currently used vacuum cups.Keywords
This publication has 9 references indexed in Scilit:
- Clinical Evaluation of a “Hand Pump” Vacuum Delivery DeviceObstetrics & Gynecology, 2002
- Clinical evaluation of a “hand pump” vacuum delivery device2002
- Early maternal and neonatal morbidity associated with operative delivery in second stage of labour: a cohort studyThe Lancet, 2001
- Iatrogenic vCJD from surgical instrumentsBMJ, 2001
- Operative vaginal delivery: clinical appraisal of a new vacuum extraction deviceAustralian and New Zealand Journal of Obstetrics and Gynaecology, 2001
- Choice and instrumental deliveryBJOG: An International Journal of Obstetrics & Gynaecology, 1996
- The obstetric vacuum extractor is the instrument of first choice for operative vaginal deliveryBJOG: An International Journal of Obstetrics & Gynaecology, 1989
- A multicentre randomized trial comparing delivery with a silicone rubber cup and rigid metal vacuum extractor cupsBJOG: An International Journal of Obstetrics & Gynaecology, 1989
- North Staffordshire/Wigan assisted delivery trialBJOG: An International Journal of Obstetrics & Gynaecology, 1989