Comparison between dynamic contour tonometry and Goldmann applanation tonometry correcting equations
Authors:
Open Access
- 23 November 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-11
Abstract
In order to investigate the reliability of correcting GAT formulas in comparison with dynamic contour tonometry (DCT), this study included 112 right eyes of 112 healthy subjects aged from 21 to 77 years, whose eyes underwent to a full ophthalmologic exam. IOP was measured in each eye with DCT and then with GAT. IOP values obtained with GAT were corrected with 10 equations and then compared with those provided by DCT. Participants mean age was 42.24 ± 14.08 years; mean IOP measured with DCT was 17.61 ± 2.87 and 15.50 ± 2.47 mmHg, measured with GAT. The mean discordance between DCT and GAT measurements was 2.11 ± 2.24 mmHg. All the correcting formulas, but Srodka one (p ˂ 0.001), tend to increase the difference between GAT and DCT. According to these results Śródka equation provides the best correction, reducing the difference between the two IOP measurement methods of − 0.03 ± 0.85 mmHg. Other equations do not provide a valid improvement of the agreement between the two methods or they provide a worsening of the agreement.Keywords
This publication has 31 references indexed in Scilit:
- Poor utility of intraocular pressure correction formulae in individual glaucoma and glaucoma suspect patientsClinical & Experimental Ophthalmology, 2011
- Multiparameter Correction Equation for Goldmann Applanation TonometryOptometry and Vision Science, 2011
- Goldmann Tonometry Correction Factors Based on Numerical AnalysisJournal of Biomechanical Engineering, 2009
- Dynamic Contour Tonometry in Comparison to Intracameral IOP MeasurementsInvestigative Ophthalmology & Visual Science, 2008
- Correlation Between Intraocular Pressure, Central Corneal Thickness, Stage of Glaucoma, and Demographic Patient DataJournal of Glaucoma, 2006
- Dynamic Contour TonometryJournal of Glaucoma, 2005
- Comparison of Dynamic Contour Tonometry with Goldmann Applanation TonometryInvestigative Ophthalmology & Visual Science, 2004
- APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESSActa Ophthalmologica, 1975
- The validity of the Imbert-Fick law as applied to applanation tonometryExperimental Eye Research, 1963
- THE SO-CALLED IMBERT-FICK LAWArchives of Ophthalmology (1950), 1960