Reducing peripheral nerve stimulation due to gradient switching using an additional uniform field coil
Authors:
Open Access
- 20 May 2011
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 66 (5) , 1498-1509
Abstract
This study shows that larger rates of change of gradient with time (dG/dt) can be achieved at the threshold for peripheral nerve stimulation by applying a uniform concomitant field varying synchronously with a transverse field gradient and that this increase may be achieved without significant reduction of the spatial extent of the region over which imaging can be carried out. Realization of similar benefits through application of a uniform, z-directed field varying synchronously with an axial gradient is also demonstrated. The design and construction of transverse and axial coil arrangements is described, along with the results of volunteer studies that were carried out on 20 subjects, with the subjects positioned with four different regions (head, heart, hips, and knees) centered in the coils. These experiments were carried out at zero-field on a prototype system in which the coils were not actively shielded. The uniform concomitant field coil was not torque balanced. The increase in the rate of change of gradient at the threshold for peripheral nerve stimulation that could be achieved by the addition of the uniform field depends on body position and was larger for the transverse coils (head = × 1.9 ± 0.6; heart = × 0.9 ± 0.3; hips = × 1.4 ± 0.4; knees = × 1.5 ± 0.4) than for the axial coils (head = × 1.5 ± 0.6; heart = × 0.8 ± 0.3; hips = × 1.3 ± 0.4; knees = × 1.1 ± 0.3). Magn Reson Med, 2011.Keywords
This publication has 18 references indexed in Scilit:
- Interaction of MRI field gradients with the human bodyPhysics in Medicine & Biology, 2009
- Magnetic resonance imaging with composite (dual) gradientsConcepts in Magnetic Resonance Part B, 2009
- Electric fields induced in the human body by time-varying magnetic field gradients in MRI: numerical calculations and correlation analysisPhysics in Medicine & Biology, 2007
- Finite difference time domain (FDTD) method for modeling the effect of switched gradients on the human body in MRIMagnetic Resonance in Medicine, 2002
- Induction of electric fields due to gradient switching: A numerical approachMagnetic Resonance in Medicine, 2002
- Gradient system providing continuously variable field characteristicsMagnetic Resonance in Medicine, 2002
- Analytic calculations of the E-fields induced by time-varying magnetic fields generated by cylindrical gradient coilsMagnetic Resonance in Medicine, 2000
- Review of Patient Safety in Time-Varying Gradient FieldsJournal of Magnetic Resonance Imaging, 2000
- Modular gradient coil: A new concept in high-performance whole-body gradient coil designMagnetic Resonance in Medicine, 1999
- Physiological Side Effects of Fast Gradient Switching1998