Design of CMOS front-end circuitry for the acquisition of biopotential signals
Authors:
- 1 December 2016
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 161-164
Abstract
This paper presents the design of the front-end circuitry of a low power and low noise analog biopotential signal acquisition system. The design has been optimized for low power consumption while still achieving the gain, bandwidth and input referred noise required by this application. The design consists of three main building circuit blocks. The first stage is the instrumentation amplifier which amplifies and filters the biopotential signals and provides a gain of 13.9 dB. This stage is based on current feedback techniques and achieves a CMRR of 124 dB, while the integrated input referred noise is 2.67 μV rms . This is followed by an analog multiplexer, which allows the system to process multiple signals. The last stage is the programmble gain amplifier, which allows the overall gain to be set according to the type of biopotential signal, in the range from 13.9 dB to 73.9 dB. Each instrumentation amplifier consumes 10 μW while the programmable gain amplifier consumes 61 μW from a 2 V supply, making the design appropriate for battery operation. A standard 0.35 μm CMOS technology was used in this work.Keywords
This publication has 8 references indexed in Scilit:
- A novel low-noise fully differential CMOS instrumentation amplifier with 1.88 noise efficiency factor for biomedical and sensor applicationsMicroelectronics Journal, 2016
- A bandwidth-tunable bioamplifier with voltage-controlled symmetric pseudo-resistorsMicroelectronics Journal, 2015
- In-vivo characterization of a 0.8 – 3 µVRMS input-noise versatile CMOS pre-amplifier2015
- The impact of device type and sizing on phase noise mechanismsIEEE Journal of Solid-State Circuits, 2005
- A CMOS IC for portable EEG acquisition systemsIEEE Transactions on Instrumentation and Measurement, 1998
- Circuit techniques for reducing the effects of op-amp imperfections: autozeroing, correlated double sampling, and chopper stabilizationProceedings of the IEEE, 1996
- AC instrumentation amplifier for bioimpedance measurementsIEEE Transactions on Biomedical Engineering, 1993
- Micropower high-performance SC building block for integrated low-level signal processingIEEE Journal of Solid-State Circuits, 1985