Abstract
Objective: To determine if an interlocking bolt would limit subsidence of the biological fixation universal hip (BFX®) femoral stem under cyclic loading and enhance construct stiffness, yield, and failure properties.Study Design: Ex vivo biomechanical study.Animals: Cadaveric canine femora (10 pairs).Methods: Paired femora implanted with a traditional stem or an interlocking stem (constructs) were cyclically loaded at walk, trot, and gallop loads while implant and bone motions were captured using kinematic markers and high‐speed video. Constructs were then loaded to failure to evaluate failure mechanical properties.Results: Implant subsidence was greater (P = .037) for the traditional implant (4.19 mm) than the interlocking implant (0.78 mm) only after gallop cyclic loading, and cumulatively after walk, trot, and gallop cyclic loads (5.20 mm vs. 1.28 mm,P = .038). Yield and failure loads were greater (P = .029 and .002, respectively) for the interlocking stem construct (1155 N and 2337 N) than the traditional stem construct (816 N and 1405 N). Version angle change after cyclic loading was greater (P = .020) for the traditional implant (3.89 degrees) than for the interlocking implant (0.16 degrees), whereas stem varus displacement at failure was greater (P = .008) for the interlocking implant (1.5 degrees) than the traditional implant (0.17 degrees).Conclusion: Addition of a stabilizing bolt enhanced construct stability and limited subsidence of a BFX®femoral stem. Use of the interlocking implant may decrease postoperative subsidence. However, in vivo effects of the interlocking bolt on osseointegration, bone remodeling, and stress shielding are unknown.Keywords
This publication has 36 references indexed in Scilit:
- Biomechanical evaluation of the helica femoral implant system using traditional and modified techniquesVeterinary Surgery, 2013
- Kinetics of jump landing in agility dogsThe Veterinary Journal, 2011
- Resistance to Subsidence of an Uncemented Femoral Stem After Cerclage Wiring of a FissureVeterinary Surgery, 2011
- Risk Factors for Femoral Fracture after Canine Press-Fit Cementless Total Hip ArthroplastyVeterinary Surgery, 2010
- Complications of cementless total hip replacementVeterinary and Comparative Orthopaedics and Traumatology, 2010
- Short‐Term Aseptic Loosening of the Femoral Component in Canine Total Hip Replacement: Effects of Cementing Technique on Cement Mantle GradeVeterinary Surgery, 2005
- Canine Uncemented Porous‐Coated Anatomic Total Hip Arthroplasty: Results of a Long‐Term Prospective Evaluation of 50 Consecutive CasesVeterinary Surgery, 1999
- Femoral Strain Distribution and Subsidence After Physiological Loading of a Cementless Canine Femoral Prosthesis: The Effects of Implant Orientation, Canal Fill, and Implant FitVeterinary Surgery, 1994
- Subsidence of an Uncemented Canine Femoral StemVeterinary Surgery, 1992
- Implantation of an Uncemented Total Hip Prosthesis Technique and Initial Results of 100 ArthroplastiesVeterinary Surgery, 1992