The mechanistic causes of peripheral intravenous catheter failure based on a parametric computational study
Open Access
- 21 February 2018
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 8 (1) , 1-12
Abstract
Peripheral intravenous catheters (PIVCs) are the most commonly used invasive medical device, yet up to 50% fail. Many pathways to failure are mechanistic and related to fluid mechanics, thus can be investigated using computational fluid dynamics (CFD). Here we used CFD to investigate typical PIVC parameters (infusion rate, catheter size, insertion angle and tip position) and report the hemodynamic environment (wall shear stress (WSS), blood damage, particle residence time and venous stasis volumes) within the vein and catheter, and show the effect of each PIVC parameter on each hemodynamic measure. Catheter infusion rate has the greatest impact on our measures, with catheter orientation also playing a significant role. In some PIVC configurations WSS was 3254 times higher than the patent vein, and blood damage was 512 times greater, when compared to control conditions. Residence time is geometry-dependent and decreases exponentially with increasing insertion angle. Stasis volume decreased with increasing infusion rate and, to a lesser degree, insertion angle. Even without infusion, the presence of the catheter changes the flow field, causing low velocity recirculation at the catheter tip. This research demonstrates how several controllable factors impact important mechanisms of PIVC failure. These data, the first of their kind, suggest limiting excessive infusion rates in PIVC.Keywords
This publication has 24 references indexed in Scilit:
- The flow field near a venous needle in hemodialysis: A computational studyHemodialysis International, 2013
- The Effect of Catheter to Vein Ratio on Blood Flow Rates in a Simulated Model of Peripherally Inserted Central Venous CathetersChest, 2011
- Effects of Disturbed Flow on Vascular Endothelium: Pathophysiological Basis and Clinical PerspectivesPhysiological Reviews, 2011
- A shear gradient–dependent platelet aggregation mechanism drives thrombus formationNature Medicine, 2009
- Peripheral venous catheters: an under-evaluated problemInternational Journal of Antimicrobial Agents, 2009
- The Incidence of PICC Line–Associated Thrombosis With and Without the Use of Prophylactic AnticoagulantsJournal of Parenteral and Enteral Nutrition, 2008
- Hemodynamic changes in the cephalic vein of patients with hemodialysis arteriovenous fistulaJournal of Clinical Ultrasound, 2007
- Fast Three‐dimensional Numerical Hemolysis ApproximationArtificial Organs, 2004
- A prospective evaluation of central venous blood flow using Doppler ultrasound in patients with a long term central venous catheterEuropean Journal of Clinical Nutrition, 1997
- Acute Vascular Endothelial Changes Associated with Increased Blood Velocity GradientsCirculation Research, 1968