N Latex FLC – new monoclonal high-performance assays for the determination of free light chain kappa and lambda

Authors:

Abstract
Background: High serum concentrations of monoclonal free light chain (FLC) kappa or lambda are markers of plasma cell dyscrasia. Methods: We developed new, latex-enhanced, specific nephelometric assays based on monoclonal antibodies for the determination of FLC kappa and lambda in serum, EDTA plasma and Li-heparin plasma for use on the Siemens BN™ systems. Results: Reference ranges were determined from 369 samples: FLC kappa 6.7–22.4 mg/L, FLC lambda 8.3–27.0 mg/L and kappa/lambda ratio 0.31–1.56. Protection from falsely low results due to antigen excess is obtained with a built-in pre-reaction in the assay protocols. Lot-to-lot consistency between three different lots of reagent, calibrators and supplementary reagent lots showed normalized differences Conclusions: N Latex FLC demonstrates high precision, good lot-to-lot consistency and freedom from a high-dose hook effect. The method comparison between Freelite™ and the N Latex FLC assays showed good clinical concordance. Further studies need to reveal the clinical value of the new FLC assays.

This publication has 20 references indexed in Scilit: