Pitfalls and solutions of the fully-automated radiosynthesis of [11C]metoclopramide
Open Access
- 18 December 2019
- journal article
- research article
- Published by Springer Nature in EJNMMI Radiopharmacy and Chemistry
- Vol. 4 (1) , 1-14
Abstract
Background: [11C]Metoclopramide is a new radiotracer for investigating the activity of P-glycoprotein at the blood-brain barrier. A highly stable and reproducible radiosynthesis is a prerequisite for clinical studies applying [11C]metoclopramide or other 11C-labelled radiotracers, therefore all potential pitfalls must be identified and monitored to allow a stable process. Results: Long-term production (n = 94 in a time range of approximately 2 years) of [11C]metoclopramide synthesized on two commercially available synthesizers yielded 3.9 ± 2.0 GBq of product with a molar activity of 132 ± 164 GBq/μmol and an overall success rate of 93%. During all successful productions, the product quality was in accordance with the recommendations of the European Pharmacopoeia. The most common pitfalls that were identified for the radiosynthesis included poor turnover into [11C]CH3OTf, decomposition of the solvent or insufficient semi-preparative HPLC performance. Conclusion: The study provides long-term insight in the improved, robust and stable preparation of [11C]metoclopramide for human use.Keywords
Funding Information
- FWF (KLI694-B30)
This publication has 14 references indexed in Scilit:
- Molar activity – The keystone in 11C-radiochemistry: An explorative study using the gas phase methodNuclear Medicine and Biology, 2018
- L-[S-methyl-11C]methionine – An example of radiosynthetic optimizationApplied Radiation and Isotopes, 2018
- Comparison of fully-automated radiosyntheses of [11C]erlotinib for preclinical and clinical use starting from in target produced [11C]CO2 or [11C]CH4EJNMMI Radiopharmacy and Chemistry, 2018
- P-Glycoprotein (ABCB1) Inhibits the Influx and Increases the Efflux of11C-Metoclopramide Across the Blood–Brain Barrier: A PET Study on Nonhuman PrimatesJournal of Nuclear Medicine, 2018
- Positron Emission Tomography Imaging Reveals an Importance of Saturable Liver Uptake Transport for the Pharmacokinetics of MetoclopramideContrast Media & Molecular Imaging, 2018
- Speed matters to raise molar radioactivity: Fast HPLC shortens the quality control of C-11 PET-tracersNuclear Medicine and Biology, 2018
- Radiosynthesis and evaluation of an 18F–labeled silicon containing exendin-4 peptide as a PET probe for imaging insulinomaEJNMMI Radiopharmacy and Chemistry, 2018
- Consensus nomenclature rules for radiopharmaceutical chemistry — Setting the record straightNuclear Medicine and Biology, 2017
- Imaging the Impact of the P-Glycoprotein (ABCB1) Function on the Brain Kinetics of MetoclopramideJournal of Nuclear Medicine, 2015
- Synthesis of [11C]iodomethane by iodination of [11C]methaneApplied Radiation and Isotopes, 1997