Abstract
We sought to present the 5-year clinical outcome of the titanium-nitride-oxide-coated bioactive stents (BAS), as compared to paclitaxel-eluting stents (PES), in a real-world patient population. From May 2003 to November 2004, we enrolled 405 consecutive patients who underwent percutaneous coronary intervention with either BAS or PES implantation. Patients were prospectively followed up for 5 years. The primary end-point was major adverse cardiac events (MACE) at 5-year follow-up including cardiac death, nonfatal myocardial infarction (MI), or target lesion revascularization. A total of 201 patients received BAS (218 lesions/221 stents) while 204 patients received PES (244 lesions/247 stents). Clinical follow-up for 5 years was completed in all patients. Cumulative MACE at the end of 5-year follow-up occurred in 34 (16.9%) patients in the BAS group, as compared to 53 (26%) in the PES group (OR 1.7, 95% CI 1.1-2.8, P = 0.03). This difference was mainly driven by a lower incidence of MI in the BAS group as compared with the PES group (9.5% vs. 20.6%, OR 2.5, 95% CI 1.4-4.4, P = 0.002). Stent thrombosis occurred in 16 (7.8%) patients in the PES group, while no one suffered stent thrombosis in the BAS group. BAS implantation in a real-world patient population achieves an excellent clinical outcome over 5-year follow-up, with a significantly lower incidence of MI, MACE, and stent thrombosis as compared to PES.