Are Virtually Designed 3D Printed Surgical Splints Accurate Enough for Maxillary Reposition as an Intermediate Orthognathic Surgical Guide
- 9 June 2023
- journal article
- research article
- Published by Springer Nature in Journal of Maxillofacial and Oral Surgery
- Vol. 22 (4) , 861-872
Abstract
Background and Aim The accuracy of the virtually-designed 3D-printed surgical splints requires investigation for the practical use of surgical plan in the operating room. This study aimed to compare the validity of the 3D-printed and the conventional intermediate splints and evaluate the outcomes after the surgical application of the 3D-printed splint compared with the predicted values. Methods In this study, ten patients with dentofacial deformity were recruited. Participants were analyzed by the conventional surgical planning and virtual surgical planning. The intermediate surgical splints were created by the conventional and 3D-printing methods. Maxillary movements in 3 spatial directions were measured in an articulator after the application of both splints. Correlation and agreement between the two methods were tested by intraclass correlation coefficient (ICC). After the confirmation of 3D printed splint validity for each patient, the surgery was performed using 3D printed splints. It is assumed that ideally cephalometric prediction values are going to be obtained using conventional acrylic splints (gold standard). So, as a second objective, the outcome of the surgically-applied 3D-printed splint was evaluated and compared with the predicted values and finally analyzed by the paired t-test. Results Based on the observations, there was an excellent agreement between the virtually-designed 3D-printed and conventional intermediate surgical splints (ICC ranged between 0.83 and 0.99 for linear values). There was a good cumulative agreement of ICC greater than 0.80. Overall, the mean linear measurements were not different between conventional and 3D-printed splint on the articulator. Also, there were no significant differences between the linear and angular measurements of 2D-cephalometric prediction and postoperation values. Conclusion The results showed cautiously the acceptable accuracy of the 3D-printed splints for several parameters in three spatial dimensions within the laboratory and clinical settings.Keywords
This publication has 25 references indexed in Scilit:
- How accurate are rapid prototyped (RP) final orthognathic surgical wafers? A pilot studyBritish Journal of Oral and Maxillofacial Surgery, 2014
- Accuracy of perioperative mandibular positions in orthognathic surgeryInternational Journal of Oral and Maxillofacial Surgery, 2014
- RapidSplint: virtual splint generation for orthognathic surgery – results of a pilot seriesComputer Aided Surgery, 2014
- A Paradigm Shift in Orthognathic Surgery? A Comparison of Navigation, Computer-Aided Designed/Computer-Aided Manufactured Splints, and “Classic” Intermaxillary Splints to Surgical Transfer of Virtual Orthognathic PlanningJournal of Oral and Maxillofacial Surgery, 2013
- Accuracy of Upper Jaw Positioning With Intermediate Splint Fabrication After Virtual Planning in Bimaxillary Orthognathic SurgeryJournal of Craniofacial Surgery, 2013
- Comparison of the accuracy of the three-dimensional virtual method and the conventional manual method for model surgery and intermediate wafer fabricationOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2009
- Manufacturing splints for orthognathic surgery using a three-dimensional printerOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2008
- Two-Dimensional Cephalometry and Computerized Orthognathic Surgical Treatment PlanningClinics in Plastic Surgery, 2007
- Quantifying Test-Retest Reliability Using the Intraclass Correlation Coefficient and the SEMJournal of Strength & Conditioning Research, 2005
- Stability and predictability of orthognathic surgeryAmerican Journal of Orthodontics and Dentofacial Orthopedics, 2004